取出的双动全髋关节置换系统的微动腐蚀损伤
本研究在机构审查委员会批准的植入物取出方案下,纳入17例患者的18套双动全髋关节置换系统,按Goldberg法对头颈交界处进行微动和腐蚀评分并宏观检查各部件损伤模式。平均植入时长13.4个月,呈现轻至中度微动腐蚀损伤,聚乙烯内衬出现边缘变形、划痕和点蚀,金属部件出现抛光擦伤和划痕。结论认为双动全髋系统可能易发生与传统模块化全髋系统相似的微动腐蚀损伤,需未来研究确认这些结果及临床意义。
要点
- 平均植入13.4个月的双动全髋系统呈现轻至中度微动腐蚀损伤。
- 聚乙烯内衬见边缘变形、划痕和点蚀,金属部件见抛光擦伤和划痕。
- 微动与腐蚀总评分强相关(ρ=0.967,P<.0001)。
核心结论
- 双动全髋关节置换系统可能易发生与传统模块化全髋关节置换系统相同的微动腐蚀损伤,需未来研究确认这些结果及临床意义。
研究设计与样本
- 在机构审查委员会批准的植入物取出方案下,纳入17例患者的18套双动全髋关节置换系统。
- 按Goldberg等的方法对头颈交界处进行微动和腐蚀评分,并对部件进行宏观损伤模式检查。
- 从病历收集人口学和手术数据,评估X线片上的部件位置。
- 数据分析采用Spearman秩相关和Mann-Whitney U检验,α=0.05。
主要发现
- 平均植入时长为13.4个月,呈现轻至中度微动腐蚀损伤。
- 聚乙烯内衬出现边缘变形、划痕和点蚀损伤,金属部件出现抛光擦伤和划痕损伤。
- 微动与腐蚀总评分强相关(ρ=0.967,P<.0001)。
- 腐蚀总评分与聚乙烯内衬嵌入的存在中度相关(ρ=0.690,P=.017)。
- 聚乙烯内衬磨损与股骨柄锥边缘变形呈中度正相关(ρ=0.690,P=.017)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 未来研究需要确认这些结果及临床意义。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Dual-mobility (DM) total hip arthroplasty (THA) systems are designed to increase stability while potentially avoiding problems associated with large femoral heads. Complications of these systems are not yet fully understood. This study aims at characterizing in vivo performance of DM hip systems and assessing modes of clinical failure. METHODS: Under an institutional review board-approved implant retrieval protocol, 18 DM THA systems from 17 patients were included. Implants were graded at the head-neck junction for fretting and corrosion based on the system of Goldberg et al. Components were also macroscopically examined for different damage modes. Demographics and surgical data were collected from medical records, and radiographs were assessed for component positioning. Data were analyzed through Spearman rank-order correlation and Mann-Whitney U-tests, with α = 0.05. RESULTS: The average length of implantation was 13.4 months with mild to moderate fretting corrosion damage. Polyethylene (PE) liners exhibited edge deformation, scratching, and pitting damage. Metallic components exhibited burnishing and scratching damage. Summed fretting and corrosion scores were strongly correlated (ρ = 0.967, P < .0001). Summed corrosion score was moderately correlated with presence of embedding on the PE liner (ρ = 0.690, P = .017). PE liner abrasion and edge deformation of the femoral stem taper were moderately positively correlated (ρ = 0.690, P = .017). Fretting and corrosion damage were not significantly correlated with patient demographics or radiographic positioning of implants. There were no differences in scores between modular and monoblock designs. CONCLUSION: These findings demonstrate that dual-mobility THA systems may be susceptible to the same fretting and corrosion damage observed in traditional modular THA systems. Future studies are needed to confirm these results and clinical significance.
引用
- Lombardo DJ, Siljander MP, Gehrke CK, et al. Fretting and Corrosion Damage of Retrieved Dual-Mobility Total Hip Arthroplasty Systems. J Arthroplasty. 2019. PMID: 30853157. doi:10.1016/j.arth.2019.02.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。