CT结肠镜检查中股骨近端Hounsfield单位与双能X线吸收测定法结果相关
本回顾性研究在五个中心2010至2017年间9085例CT结肠镜检查中纳入277例有DEXA结果者,测量股骨近端Hounsfield单位(HU)并与DEXA结果比较。综合HU与单次HU测量均与股骨近端T值呈正相关(r分别为0.75和0.74),单次HU模型识别低骨密度的AUC为0.82。作者认为该技术可重复测量股骨近端HU并预测低骨密度,未来需在更大研究中验证。
要点
- 综合HU与股骨近端T值相关系数0.75(95%CI 0.69-0.80),单次HU为0.74(95%CI 0.68-0.79)。
- 单次HU模型识别低骨密度的AUC为0.82(95%CI 0.77-0.87)。
- 阈值113 HU识别低骨密度的敏感度73%、特异度76%。
核心结论
- 该测量技术可在CT结肠镜检查中可重复地测量股骨近端HU,基于此技术的股骨近端HU可预测低骨密度,这些CT扫描常在首次DEXA之前进行,可能有助于更早识别低骨密度。
- 未来需在更大研究中验证这些结果,并确定这些结果能否预判未来骨折风险。
研究设计与样本
- 回顾性研究,2010至2017年间五个中心共行9085例CT结肠镜检查,其中277例(3%)有可用DEXA结果并纳入,8809例(97%)因CT成像不充分、缺乏DEXA筛查或缺乏股骨近端DEXA结果被排除。
- CT结肠镜检查与DEXA扫描之间中位天数为595天,未因扫描间隔时间排除患者;两名评估者在CT结肠镜检查图像上股骨近端四个点测量HU,采用组内相关系数评估评估者间信度。
- 采用Pearson相关系数比较综合HU(八次测量平均值)和单次HU测量与每项DEXA结果(股骨近端BMD、股骨近端T值、股骨颈BMD、股骨颈T值);根据最低DEXA T值将患者分层为骨质疏松、骨量减少或骨密度正常;计算受试者工作特征曲线下面积(AUC)评估单次HU值识别低骨密度可能阈值的分类能力。
- 各亚组分析计算DEXA与HU之间的Pearson相关系数,并在线性回归模型中使用交互检验评估各亚组对整体预测模型的贡献。
- 证据等级为III级,诊断性研究。
主要发现
- 综合HU和单次HU测量与股骨近端T值的Pearson相关系数最高,分别为0.75(95%CI 0.69-0.80)和0.74(95%CI 0.68-0.79)。
- 综合HU和单次HU测量的评估者间信度(组内相关系数)分别为0.97(95%CI 0.72-0.99)和0.96(95%CI 0.89-0.98)。
- 根据DEXA结果,20例为骨质疏松,167例为骨量减少,90例骨密度正常;骨质疏松患者综合HU均值为70±30 HU,骨量减少患者为110±36 HU,骨密度正常患者为158±43 HU(p<0.001)。
- 单次HU模型的AUC为0.82(95%CI 0.77-0.87);阈值214 HU识别低骨密度的敏感度为100%,阈值59 HU的特异度为100%;阈值113 HU的敏感度为73%、特异度为76%。
- 按十年年龄组分层,各十年年龄组综合HU与股骨近端T值均呈正相关,范围0.71至0.83(95%CI 0.59-0.91);按CT与DEXA间隔>6个月或<6个月分层(0.75;95%CI 0.62-0.84)以及按性别分层(0.70-0.76;95%CI 0.48-0.81)亦呈正相关;线性回归模型显示HU与股骨近端T值的整体正相关系数不受任何亚组影响。
关键数值与推荐
- 综合HU与股骨近端T值相关系数0.75(95%CI 0.69-0.80),单次HU为0.74(95%CI 0.68-0.79)。
- 综合HU评估者间信度0.97(95%CI 0.72-0.99),单次HU为0.96(95%CI 0.89-0.98)。
- 骨质疏松患者综合HU均值70±30 HU,骨量减少患者110±36 HU,骨密度正常患者158±43 HU(p<0.001)。
- 单次HU模型AUC为0.82(95%CI 0.77-0.87);阈值214 HU敏感度100%,阈值59 HU特异度100%,阈值113 HU敏感度73%、特异度76%。
临床怎么用
- 原文未据此给出推荐。
局限性
- 未来需在更大研究中验证这些结果,并确定这些结果能否预判未来骨折风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Quantifying bone mineral density (BMD) on CT using commercial software demonstrates good-to-excellent correlations with dual-energy x-ray absorptiometry (DEXA) results. However, previous techniques to measure Hounsfield units (HUs) within the proximal femur demonstrate less successful correlation with DEXA results. An effective method of measuring HUs of the proximal femur from CT colonoscopy might allow for opportunistic osteoporosis screening. QUESTIONS/PURPOSES: (1) Do proximal femur HU measurements from CT colonoscopy correlate with proximal femur DEXA results? (2) How effective is our single HU measurement technique in estimating the likelihood of overall low BMD? (3) Does the relationship between our comprehensive HU measurement and DEXA results change based on age, sex, or time between studies? METHODS: This retrospective study investigated the measurement of HU of the femur obtained on CT colonoscopy studies compared with DEXA results. Between 2010 and 2017, five centers performed 9085 CT colonoscopy studies; of those, 277 (3%) also had available DEXA results and were included in this study, whereas 8809 (97%) were excluded for inadequate CT imaging, lack of DEXA screening, or lack of proximal femur DEXA results. The median number of days between CT colonoscopy and DEXA scan was 595 days; no patient was excluded based on time between scans because bone remodeling is a long-term process and this allowed subgroup analysis based on time between scans. Two reviewers performed HU measurements at four points within the proximal femur on the CT colonoscopy imaging and intraclass correlation coefficients were used to evaluate interrater reliability. We used Pearson correlation coefficients to compare the comprehensive (average of eight measurements) and a single HU measurement with each DEXA result-proximal femur BMD, proximal femur T-score, femoral neck BMD, and femoral neck T-score-to identify the best measurement technique within this study. Based on their lowest DEXA T-score, we stratified patients to a diagnosis of osteoporosis, osteopenia, or normal BMD. We then calculated the area under the receiver operator characteristic curves (AUCs) to evaluate the classification ability of a single HU value to identify possible threshold(s) for detecting low BMD. For each subgroup analysis, we calculated Pearson correlation coefficients between DEXA and HUs and evaluated each subgroup's contribution to the overall predictive model using an interaction test in a linear regression model. RESULTS: The Pearson correlation coefficient between both the comprehensive and single HU measurements was highest compared with the proximal femur T-score at 0.75 (95% confidence interval [CI], 0.69-0.80) and 0.74 (95% CI, 0.68-0.79), respectively. Interobserver reliability, measured with intraclass correlation coefficients, for the comprehensive and single HU measurements was 0.97 (95% CI, 0.72-0.99) and 0.96 (95% CI, 0.89-0.98), respectively. Based on DEXA results, 20 patients were osteoporotic, 167 had osteopenia, and 90 patients had normal BMD. The mean comprehensive HU for patients with osteoporosis was 70 ± 30 HUs; for patients with osteopenia, it was 110 ± 36 HUs; and for patients with normal BMD, it was 158 ± 43 HUs (p < 0.001). The AUC of the single HU model was 0.82 (95% CI, 0.77-0.87). A threshold of 214 HUs is 100% sensitive and 59 HUs is 100% specific to identify low BMD; a threshold of 113 HUs provided 73% sensitivity and 76% specificity. When stratified by decade age groups, each decade age group demonstrated a positive correlation between the comprehensive HU and proximal femur T-score, ranging between 0.71 and 0.83 (95% CI, 0.59-0.91). Further subgroup analysis similarly demonstrated a positive correlation between the comprehensive HU and proximal femur T-score when stratified by > 6 months or < 6 months between CT and DEXA (0.75; 95% CI, 0.62-0.84) as well as when stratified by sex (0.70-0.76; 95% CI, 0.48-0.81). The linear regression model demonstrated that the overall positive correlation coefficient between HUs and the proximal femur T-score is not influenced by any subgroup. CONCLUSIONS: Our measurement technique provides a reproducible measurement of HUs within the proximal femur HUs on CT colonoscopy. Hounsfield units of the proximal femur based on this technique can predict low BMD. These CT scans are frequently performed before initial DEXA scans are done and therefore may lead to earlier recognition of low BMD. Future research is needed to validate these results in larger studies and to determine if these results can anticipate future fracture risk. LEVEL OF EVIDENCE: Level III, diagnostic study.