每次后续更换约束衬垫在预防不稳定方面的效果递减
该研究基于1989年至2016年本机构全关节登记系统,识别554例首次更换约束衬垫,其中40例机械失败后更换为第二枚约束衬垫,13例再次机械失败后更换为第三枚约束衬垫。首次、第二次和第三次约束衬垫4年无机械失败翻修生存率分别为90%、52%和28%。与首次约束衬垫相比,第二枚和第三枚约束衬垫因机械失败翻修的比值比分别为8(P<.0001)和10(P<.0001)。
要点
- 首次、第二次和第三次约束衬垫4年无机械失败翻修生存率分别为90%、52%和28%。
- 第二枚约束衬垫因机械失败翻修的比值比为8(P<.0001),第三枚为10(P<.0001)。
- 更换为不同设计的约束衬垫并未降低失败风险。
核心结论
- 更换为第二枚或第三枚约束衬垫后机械失败率高。
- 当全髋关节置换在约束衬垫后出现不稳定时,术者在更换为另一枚约束衬垫前应谨慎。
研究设计与样本
- 回顾性研究,利用本机构全关节登记系统,1989年至2016年识别554例首次更换约束衬垫。
- 554例中40例因初次约束衬垫机械失败更换为第二枚约束衬垫,40例中13例因第二枚约束衬垫机械失败更换为第三枚约束衬垫。
- 纳入失败分析的患者至少随访2年,平均随访5.5年。
主要发现
- 首次、第二次和第三次约束衬垫4年无机械失败翻修生存率分别为90%、52%和28%。
- 与首次约束衬垫相比,第二枚约束衬垫因机械失败翻修的比值比为8(P<.0001)。
- 与首次约束衬垫相比,第三枚约束衬垫因机械失败翻修的比值比为10(P<.0001)。
- 使用不同约束衬垫未降低失败风险。
临床怎么用
- 当全髋关节置换在约束衬垫后出现不稳定时,作者认为术者在更换为另一枚约束衬垫前应谨慎。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Constrained liners are used to treat recurring total hip arthroplasty dislocations in certain patients. Little is known about the fate of constrained liner revisions. The aim of this study is to evaluate the rate and survivorship free of revision for mechanical failure of constrained liners in patients who were revised to their first, second, and third constrained liners. METHODS: From 1989 to 2016, using our institution's total joint registry, we identified 554 revisions to a first constrained liner. Of the 554, 40 had mechanical failure of their index constrained liner and had revision to a second constrained liner. Of the 40, 13 had mechanical failure of their second constrained liner and had revision to a third constrained liner. Patients included in the failure analysis had minimum 2-year follow-up, with a mean follow-up of 5.5 years. RESULTS: In patients receiving their first, second, and third constrained liners, the survivorship free of revision for mechanical failure at 4 years was 90%, 52%, and 28%, respectively. Patients with a second and third constrained liner were more likely to have a revision for mechanical failure (second odds ratio 8, P < .0001; third odds ratio 10, P < .0001) compared to those receiving their first constrained liner. Using a different constrained liner did not decrease the risk of failure. CONCLUSION: After revision to a second or third constrained liner there is high rate of mechanical failure. When a total hip arthroplasty becomes unstable after constrained liner, surgeons should exercise caution before revising to another constrained liner.
引用
- Hernandez NM, Sierra RJ, Trousdale RT, et al. Constrained Liner Revision Is Less Effective With Each Subsequent Constrained Liner Revision at Preventing Instability. J Arthroplasty. 2019. PMID: 30773358. doi:10.1016/j.arth.2019.01.061
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。