放射立体测量分析可体内测量全膝关节置换术中极小磨损量

Radiostereometric Analysis Permits In Vivo Measurement of Very Small Levels of Wear in TKA.
PubMed 文献 · Clin Orthop Relat Res · 2019-01-01

该回顾性研究纳入49例接受同一后稳定型钴铬对传统聚乙烯全膝关节置换假体、随访至少10年的患者,采用放射立体测量分析(RSA)在单一时间点测量线性磨损率。结果显示最大线性磨损率外侧为0.047 mm/年、内侧为0.052 mm/年;男性磨损率在内侧和外侧均高于女性;磨损与身高增加、体重指数降低、假体尺寸增大及下肢内翻 alignment 相关。

要点

  • 最大线性磨损率外侧0.047 mm/年,内侧0.052 mm/年。
  • 男性磨损率内侧比女性高0.022 mm/年(p<0.001),外侧高0.012 mm/年(p=0.008)。
  • 下肢内翻程度越大,内侧磨损越大(ρ=0.33,p=0.02)。

核心结论

  • 磨损率较高与人口学因素和下肢力线相关。
  • 对其他现代假体系统进行进一步RSA磨损研究可为取回研究提供补充信息及有价值的耐磨性数据。

研究设计与样本

  • 回顾性选择49例患者,随访至少10年(中位12年,范围10-20年)。
  • 研究期间共实施4082例全膝关节置换术,其中2085例为本研究检查的假体。
  • 71例符合标准(包括术后有可用全长腿X线片),其中34例返回检查,另15例来自另一项RSA研究且符合标准。
  • 所有患者接受同一假体系统的后稳定型钴铬对传统聚乙烯全膝关节假体,该假体为当时本机构最常用。
  • 患者在单一时间点于0°至120°屈曲范围接受站立位RSA检查,不使用标记珠。

主要发现

  • 采用任一屈曲角度的最大线性磨损率,外侧为0.047 mm/年(IQR 0.034-0.066 mm/年),内侧为0.052 mm/年(IQR 0.040-0.069 mm/年)。
  • 采用所有屈曲角度线性磨损率的中位数,外侧为0.027 mm/年(IQR 0.017-0.046 mm/年),内侧为0.038 mm/年(IQR 0.022-0.054 mm/年)。
  • 男性内侧磨损率为0.049 mm/年(IQR 0.042-0.077 mm/年),外侧为0.032 mm/年(IQR 0.026-0.059 mm/年);女性内侧为0.027 mm/年(IQR 0.016-0.039 mm/年),外侧为0.020 mm/年(IQR 0.013-0.032 mm/年)。
  • 男性磨损率在内侧大于女性(差值=0.022 mm/年,p<0.001),在外侧也大于女性(差值=0.012 mm/年,p=0.008)。
  • 磨损增加与身高增加相关(内侧ρ=0.48,p<0.001;外侧ρ=0.30,p=0.04),与体重指数降低相关(内侧ρ=-0.31,p=0.03),与假体尺寸增大相关(内侧ρ=0.34,p=0.02)。

临床怎么用

  • 作者认为,长期随访中表现良好的假体展示了良好的耐磨性,磨损量与取回研究报告值一致。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Revision of TKA as a result of polyethylene wear is decreasing, but long-term wear performance of polyethylene is still a topic of interest to surgeons and device manufacturers seeking to improve longevity. Measuring wear of modern, wear-resistant implants has been described using radiostereometric analysis (RSA). Performing in vivo measurements would establish whether implant retrieval studies are representative of wear in well-performing knees. QUESTIONS/PURPOSES: For a single knee implant system, we sought to determine (1) the linear wear rate using RSA; (2) the association between demographic factors and wear rate; and (3) the association between limb alignment and wear rate. METHODS: A total of 49 patients with a minimum followup of 10 years (median, 12 years; range, 10-20 years) were retrospectively selected. During the examined period, 4082 TKAs were performed of which 2085 were the implant examined in this study. There were 71 of these patients who met the criteria including an available full-leg radiograph postoperatively, and 34 of these patients returned for examination along with 15 additional from a separate RSA study that also met the criteria. All patients received a posterior-stabilized, cobalt-chromium-on-conventional polyethylene total knee implant from a single implant system, which was the most commonly used at our institution at the time. Patients underwent standing RSA examinations from 0° to 120° of flexion at a single time point without the use of marker beads. Linear wear rates (including creep) were measured based on intersections between the femoral component and tibial insert models. Associations between wear and patient age at surgery, sex, height, weight, body mass index, tibial insert size, and limb alignment were examined. RESULTS: Using the maximum linear wear rate from any flexion angle, the lateral rate was 0.047 mm/year (interquartile range [IQR], 0.034-0.066 mm/year) and the medial rate was 0.052 mm/year (IQR, 0.040-0.069 mm/year). Using the median of the linear wear rates across all flexion angles, the lateral rate was 0.027 mm/year (IQR, 0.017-0.046 mm/year) and the medial rate was 0.038 mm/year (IQR, 0.022-0.054 mm/year). This rate for males was 0.049 mm/year medially (IQR, 0.042-0.077 mm/year) and 0.032 mm/year laterally (IQR, 0.026-0.059 mm/year), and for females was 0.027 mm/year medially (0.016-0.039 mm/year) and 0.020 mm/year laterally (IQR, 0.013-0.032 mm/year). The wear rate for males was greater medially (difference = 0.022 mm/year, p < 0.001) and laterally (difference = 0.012 mm/year, p = 0.008). There were associations between greater wear and increasing height (ρ = 0.48, p < 0.001 medially and ρ = 0.30, p = 0.04 laterally), decreasing body mass index (ρ = -0.31, p = 0.03 medially), and greater implant size (ρ = 0.34, p = 0.02 medially). Increasingly varus leg alignment was associated with greater medial wear (ρ = 0.33, p = 0.02). CONCLUSIONS: Greater wear rates were associated with demographic factors and leg alignment. Further RSA wear studies of other modern implant systems would provide complementary information to retrieval studies and valuable data on wear resistance. CLINICAL RELEVANCE: Good wear resistance was demonstrated by well-performing implants in patients at long-term followup with wear magnitudes in agreement with reported values from retrieval studies.