人工老化与取出的氧化锆增韧氧化铝股骨头的相变与粗糙化
研究评估未植入的氧化锆增韧氧化铝(ZTA)股骨头在严重人工老化后的相变与表面粗糙化,并与取出的ZTA股骨头进行临床相关性比较;Y-TZP作为阳性对照组,Mg-PSZ作为稳定对照组。结果显示ZTA经48小时水热老化后单斜相浓度从12.2%升至21.3%,表面出现不规则突起且粗糙度增加;取出的ZTA股骨头在体内2.7年后单斜相达33%,人工老化测试低估了体内相变程度。
要点
- ZTA股骨头经48小时水热老化后单斜相浓度从12.2%升至21.3%,表面粗糙度显著增加。
- 取出的ZTA股骨头在体内2.7年后单斜相浓度为33%,人工老化测试低估了体内相变程度。
- ZTA股骨头的相变与表面粗糙化持续增加且未达平台期,可能导致应力集中和陶瓷材料弱化。
核心结论
- 基于ZTA取出物数据,人工老化测试低估了体内相变程度。
- ZTA股骨头的相变与表面粗糙化持续增加且未达平台期,可能导致应力集中和陶瓷材料弱化,并可能引起晚期骨折和磨损。
研究设计与样本
- 研究评估未植入的ZTA、Y-TZP和Mg-PSZ股骨头在2轮24小时水热老化前后的单斜相浓度、表面形貌和粗糙度,采用X射线衍射和光学轮廓术测量;取出的ZTA股骨头用相同方法表征。
主要发现
- ZTA股骨头经48小时高压灭菌后表面出现不规则间距的突起,高约20-30 nm、直径约100-150 μm,单斜相浓度从12.2%升至21.3%,表面粗糙度显著增加。
- 取出的ZTA股骨头表面观察到类似特征,体内2.7年后单斜相浓度为33%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Zirconia-toughened alumina (ZTA) used in hip arthroplasty contains yttria-stabilized zirconia (Y-TZP) as a toughening agent. However, Y-TZP is well known to degrade in vivo from tetragonal to monoclinic phase transformation. The stability of never-implanted ZTA femoral heads was evaluated in a severe artificial aging test, with retrieved ZTA heads also evaluated for clinical relevance. We hypothesized that ZTA would degrade due to tetragonal-to-monoclinic phase transformation, with changes in surface topography and progressive roughening. Y-TZP specimens served as a positive comparison group, while magnesia-stabilized zirconia (Mg-PSZ), which does not undergo phase transformation, served as a stable comparison group. METHODS: Monoclinic phase concentration, surface topography, and roughness of never-implanted ZTA, Y-TZP, and Mg-PSZ heads were measured by X-ray diffraction and optical profilometry, before and after 2 rounds of 24 hours of hydrothermal aging. Explanted ZTA heads were characterized by the same methods. RESULTS: After 48 hours in an autoclave, the surface of ZTA heads exhibited irregularly spaced protrusions about 20-30 nm high by 100-150 μm in diameter, with significant increases in monoclinic phase concentration (from 12.2% to 21.3%) and surface roughness. Similar features were observed on the surface of explanted ZTA heads, with 33% monoclinic phase after 2.7 years in vivo. CONCLUSION: Based on data collected from ZTA retrievals, this artificial aging test underestimated the amount of phase transformation in vivo. Phase transformation and surface roughening of ZTA heads steadily increased without reaching a plateau, which may lead to stress concentrations and weakening of the ceramic material, and could result in late fracture and wear.
引用
- Roy ME, Noel OF, Whiteside LA, et al. Phase Transformation and Roughening in Artificially Aged and Retrieved Zirconia-Toughened Alumina Femoral Heads. J Arthroplasty. 2018. PMID: 30642707. doi:10.1016/j.arth.2018.12.025
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。