五年时维生素E扩散高交联聚乙烯体内磨损评估:一项多中心放射立体分析研究

Evaluation of in vivo wear of vitamin E-diffused highly crosslinked polyethylene at five years: a multicentre radiostereometric analysis study.
PubMed 文献 · Bone Joint J · 2018-01-01

研究比较了维生素E扩散高交联聚乙烯(VEPE)与中度交联机械退火超高分子量聚乙烯(ModXLPE)在初次全髋关节置换术后五年的磨损性能,并评估VEPE患者的临床结果。208例患者(221髋)来自四个国际中心,193髋(87%)完成五年随访。五年时ModXLPE组股骨头穿透显著多于VEPE组(p < 0.001),两种内衬磨损率均很低,未导致骨溶解或无菌性松动所致的植入物失败。

要点

  • 五年随访时ModXLPE组股骨头穿透显著多于VEPE组(p < 0.001)。
  • ModXLPE内衬(β = 0.22,p = 0.010)和金属股骨头(β = 0.21,p = 0.013)是磨损增加的独立预测因素。
  • 磨损增加与放射透亮线出现(p = 0.866)或患者报告结局指标无关联,无患者出现骨溶解。

核心结论

  • 术后五年,VEPE(vs ModXLPE)和陶瓷(vs 金属)股骨头的患者磨损减少,两种内衬的磨损率均很低,未导致骨溶解或无菌性松动所致的植入物失败。

研究设计与样本

  • 前瞻性多中心研究,纳入四个国际中心的208例患者(221髋),采用放射立体分析(RSA)和临床结果评估,193髋(87%)完成五年随访。

主要发现

  • 术后一年 bedding in 和术后两年磨损在两种内衬间相似。
  • 术后五年ModXLPE组股骨头穿透显著多于VEPE组(p < 0.001)。
  • 磨损增加的独立预测因素为ModXLPE(vs VEPE)内衬(β = 0.22,p = 0.010)和金属(vs 陶瓷)股骨头(β = 0.21,p = 0.013)。
  • 磨损增加与放射透亮线出现(p = 0.866)或患者报告结局指标无关联。
  • 无患者有骨溶解证据。

关键数值与推荐

  • 五年随访时ModXLPE组股骨头穿透显著多于VEPE组(p < 0.001)。
  • ModXLPE内衬(β = 0.22,p = 0.010)和金属股骨头(β = 0.21,p = 0.013)是磨损增加的独立预测因素。
  • 磨损增加与放射透亮线无关联(p = 0.866),与患者报告结局指标无关联。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The primary aim of this study was to compare the wear properties of vitamin E-diffused, highly crosslinked polyethylene (VEPE) and one formulation of moderately crosslinked and mechanically annealed ultra-high molecular weight polyethylene (ModXLPE) in patients five years after primary total hip arthroplasty (THA). The secondary aim was to assess the clinical results of patients treated with VEPE by evaluating patient-reported outcome measures (PROMs), radiological evidence of fixation, and the incidence of mechanical failure. PATIENTS AND METHODS: A total of 208 patients (221 THAs) from four international centres were recruited into a prospective study involving radiostereometric analysis (RSA) and the assessment of clinical outcomes. A total of 193 hips (87%) were reviewed at the five-year follow-up. Of these, 136 (70%) received VEPE (vs ModXLPE) liners and 68 (35%) received ceramic (vs metal) femoral heads. PROMs and radiographs were collected preoperatively and at one, two, and five years postoperatively. In addition, RSA images were collected to measure PE wear postoperatively and at one, two, and five years after surgery. RESULTS: We observed similar bedding in one year postoperatively and wear two years postoperatively between the two types of liner. However, there was significantly more penetration of the femoral head in the ModXLPE cohort compared with the VEPE cohort five years postoperatively (p < 0.001). The only variables independently predictive of increased wear were ModXLPE (vs VEPE) liner type (β = 0.22, p = 0.010) and metal (vs ceramic) femoral head (β = 0.21, p = 0.013). There was no association between increased wear and the development of radiolucency (p = 0.866) or PROMs. No patient had evidence of osteolysis. CONCLUSION: Five years postoperatively, patients with VEPE (vs ModXLPE) and ceramic (vs metal) femoral heads had decreased wear. The rates of wear for both liners were very low and have not led to any osteolysis or implant failure due to aseptic loosening.