文献速递关节外科

使用放射立体分析评估新型膝关节置换系统胫骨组件的稳定性

The assessment of the stability of the tibial component of a novel knee arthroplasty system using radiostereometric analysis.
PubMed 文献 · Bone Joint J · 2018-01-01

本研究纳入30例接受新型后稳定固定平台(ATTUNE)骨水泥型全膝关节置换术的患者,采用放射立体分析(RSA)评估胫骨组件在术后6、12、24个月相对于6周基线的稳定性。术后12至24个月胫骨组件的平均最大总点运动(MTPM)为0.08 mm(标准差0.08),低于已发表的0.2 mm阈值(p < 0.001)。结论认为该新型设计胫骨组件在术后12至24个月间表现稳定,提示因无菌性松动导致中长期失败的风险可接受地低。

要点

  • 30例患者接受新型后稳定固定平台(ATTUNE)骨水泥型全膝关节置换术,男女各15例。
  • 术后12至24个月胫骨组件平均MTPM为0.08 mm(标准差0.08),低于0.2 mm阈值(p < 0.001)。
  • 患者报告结局指标持续改善。

核心结论

  • 该新型全膝关节置换系统胫骨组件在术后12至24个月评估间表现稳定,提示因无菌性松动导致中长期失败的风险可接受地低。

研究设计与样本

  • 队列研究,纳入30例接受初次骨水泥型后稳定固定平台(ATTUNE)全膝关节置换术的患者,男女各15例,手术时平均年龄64岁(标准差8),平均体重指数35.4 kg/m²(标准差7.9)。
  • 采用放射立体分析(RSA)评估胫骨组件在术后6、12、24个月相对于6周基线的稳定性,并评估患者报告结局指标。

主要发现

  • 术后12至24个月胫骨组件的平均最大总点运动(MTPM)为0.08 mm(标准差0.08),低于已发表的0.2 mm阈值(p < 0.001)。
  • 患者报告结局指标持续改善。

关键数值与推荐

  • RSA评估的胫骨组件稳定性以平均MTPM 0.08 mm(标准差0.08)低于已发表阈值0.2 mm(p < 0.001)为判断依据。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The introduction of a novel design of total knee arthroplasty (TKA) must achieve outcomes at least as good as existing designs. A novel design of TKA with a reducing radius of the femoral component and a modified cam-post articulation has been released and requires assessment of the fixation to bone. Radiostereometric analysis (RSA) of the components within the first two postoperative years has been shown to be predictive of medium- to long-term fixation. The aim of this study was to assess the stability of the tibial component of this system during this period of time using RSA. PATIENTS AND METHODS: A cohort of 30 patients underwent primary, cemented TKA using the novel posterior stabilized fixed-bearing (ATTUNE) design. There was an even distribution of men and women (15:15). The mean age of the patients was 64 years (sd 8) at the time of surgery; their mean body mass index (BMI) was 35.4 kg/m 2 (sd 7.9). RSA was used to assess the stability of the tibial component at 6, 12, and 24 months compared with a six-week baseline examination. Patient-reported outcome measures were also assessed. RESULTS: The mean maximum total point motion (MTPM) of the tibial component between 12 and 24 months postoperatively was 0.08 mm (sd 0.08), which is well below the published threshold of 0.2 mm (p < 0.001). Patient-reported outcome measures consistently improved. CONCLUSION: The tibial component of this novel design of TKA showed stability between assessment 12 and 24 months postoperatively, suggesting an acceptably low risk of medium- to long-term failure due to aseptic loosening.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Turgeon TR, Gascoyne TC, Laende EK, et al. The assessment of the stability of the tibial component of a novel knee arthroplasty system using radiostereometric analysis. Bone Joint J. 2018. PMID: 30499327. doi:10.1302/0301-620X.100B12.BJJ-2018-0566.R1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。