无骨水泥全髋关节置换术中股骨锉敲击音变化的临床相关性:尺寸正确的股骨锉产生与骨几何形态直接相关的独特音频模式
本研究对105例使用Corail柄的无骨水泥全髋关节置换术中的股骨锉敲击音进行录音,排除4例后分析101例记录。结果显示,最终股骨锉敲击时出现的1 kHz范围低频声带是植入物柄尺寸合适的强预测因子,该频率与股骨长度相关,支持声音源于骨本身的假说。
要点
- 最终股骨锉敲击时出现的1 kHz范围低频声带是尺寸合适植入物柄的强预测因子。
- 该低频声带的频率与股骨长度相关,支持声音源于骨本身的假说。
- 股骨锉敲击产生的低频声可通过频谱图监测,其频率可由股骨长度预测。
核心结论
- 股骨锉敲击时产生的低频声可通过频谱图监测,其频率可由股骨长度预测,并且是植入物柄尺寸合适的良好预测因子。
研究设计与样本
- 记录105例使用Corail柄的无骨水泥全髋关节置换术中的股骨锉敲击音,排除4例后分析101例记录;36例男性、65例女性,平均年龄69.9岁(标准差12.3),体重指数中位数29 kg/m²(四分位距26至32)。
主要发现
- 最终股骨锉敲击时出现的1 kHz范围低频声带是尺寸合适植入物柄的强预测因子。
- 该低频声带的频率与股骨长度相关,支持声音源于骨本身的假说。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: Cementless femoral stems must be correctly sized and well-seated to obtain satisfactory biological fixation. The change in sound that occurs during impaction of the femoral broach is said to indicate good fit, but this has not been widely studied. We set out to find whether the presence or absence of these sound changes could predict correct sizing. PATIENTS AND METHODS: We recorded the sound generated during femoral broaching for 105 cementless total hip arthroplasties using the Corail stem. Four cases were excluded, leaving 101 recordings for analysis. There were 36 male patients and 65 female patients, with a mean age of 69.9 years (sd 12.3) and median body mass index (BMI) of 29 kg/m 2 (interquartile range (IQR) 26 to 32). The recordings were analyzed to identify the frequencies of the sounds produced during impaction of the femoral broach. RESULTS: The emergence of a low-frequency band of sound in the 1 kHz range, during the final femoral broaching, was a strong predictor of a well-sized implant stem. The frequency was related to femoral length, supporting our hypothesis that the sound arose from the bone itself. CONCLUSION: The low-frequency sound generated during femoral broaching can be monitored spectrographically, its frequency can be predicted from femoral length, and it is a good predictor of appropriate stem sizing.
引用
- McConnell JS, Saunders PRJ, Young SK, et al. The clinical relevance of sound changes produced during cementless hip arthroplasty: a correctly sized femoral broach creates a distinctive pattern of audio frequencies directly related to bone geometry. Bone Joint J. 2018. PMID: 30499313. doi:10.1302/0301-620X.100B12.BJJ-2018-0368.R2