全髋关节置换术后MRI与植入物磨损及组织反应相关:一项横断面研究
该前瞻性研究纳入181例患者187髋,在全髋关节置换翻修术前行MRI,并与取出的假体组件直接磨损、腐蚀、微动评估及组织学半定量分析对比。结果显示MRI滑膜厚度与股骨柄头颈锥部微动腐蚀损伤严重度相关(ρ=0.26),金属对金属承重面中ALTR分级与体积磨损高度相关(ρ=0.93);MRI可区分不同承重类型对应的滑膜反应。
要点
- MRI滑膜增厚与模块化股骨柄女性头颈锥部微动腐蚀损伤严重度相关(ρ=0.26,95%CI 0.12-0.39,p=0.015,n=185)。
- 金属对金属承重面中,ALTR分级与体积磨损高度相关(ρ=0.93,95%CI 0.72-0.98,p<0.001,n=10)。
- MRI上金属对聚乙烯髋表现为独特的聚合物性滑膜反应,而髋表面置换、金属对金属及模块化颈髋更常见ALTR。
核心结论
- 全髋关节置换术后MRI可区分反映植入物承重类型的滑膜反应,并与植入物磨损、腐蚀、微动的直接测量及假体周围组织磨损颗粒的组织学评估相关。
研究设计与样本
- 前瞻性研究,纳入2013年10月至2017年6月期间接受翻修手术的全髋关节置换患者181例187髋,承重类型包括金属对金属、髋表面置换、金属对聚乙烯、陶瓷对聚乙烯、陶瓷对陶瓷或模块化颈设计,术前行MRI。
- 由一名对承重面设盲的读者评估滑膜反应(Gwet‘s AC1为0.65-0.97),并与组织样本半定量组织学(Gwet’s AC1为0.92)及取出的假体组件采用Goldberg评分的半定量磨损、腐蚀和微动分析(Gwet‘s AC1为0.60-0.79)对比,同时直接非接触测量植入物磨损。
- 采用相关分析和方差分析评估指标间关联及不同植入物类型的差异。
主要发现
- MRI滑膜厚度与模块化设计中股骨柄女性头颈锥部微动和腐蚀损伤的严重度相关(ρ=0.26,95%CI 0.12-0.39,p=0.015,n=185)。
- 金属对金属承重面中ALTR分级与体积磨损相关(ρ=0.93,95%CI 0.72-0.98,p<0.001,n=10)。
- 在诊断为无菌性淋巴细胞为主型血管炎相关病变和弥漫性组织坏死的患者中,MRI滑膜厚度最高。
- MRI上金属对聚乙烯髋表现为独特的聚合物性滑膜反应,而髋表面置换、金属对金属和模块化髋更常见ALTR。
- MRI上分类为聚合物性滑膜反应的髋,其组织样本中存在的颗粒数量更多。
关键数值与推荐
- MRI滑膜厚度与股骨柄女性头颈锥部微动和腐蚀损伤严重度相关:ρ=0.26(95%CI 0.12-0.39),p=0.015,n=185。
- 金属对金属承重面中ALTR分级与体积磨损相关:ρ=0.93(95%CI 0.72-0.98),p<0.001,n=10。
- 证据等级为III级,诊断性研究。
临床怎么用
- 作者认为,全髋关节置换术后MRI可无创直接显示宿主生成的滑膜反应,有助于评估疼痛性关节置换患者的不适原因,特别是提示需更及时手术干预的滑膜反应。
- 作者认为,未来研究可从对不同植入物进行前瞻性评估中获益,以评估关节置换并发症的自然纵向病史,包括不同承重结构中ALTR的发生和患病率。
局限性
- 原文在结论中提出未来研究可从对不同植入物进行前瞻性评估中获益,以评估关节置换并发症的自然纵向病史,包括不同承重结构中ALTR的发生和患病率。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: MRI is predictive of adverse local tissue reactions (ALTRs) after THA but how MRI directly relates to implant surface wear, fretting, and trunnion corrosion at different articulations between implant components remains unclear. MRI generates high-contrast images to display soft tissues around arthroplasty and may provide a surgeon the means to distinguish and differentiate host-related synovial patterns as a response to either polyethylene wear or metal wear and corrosion products. QUESTIONS/PURPOSES: The purposes of this study were (1) to correlate findings from MRI in patients who have undergone THA with direct assessment of implant wear, corrosion, and fretting from retrieved components; and (2) to distinguish the unique synovial responses on MRI in patients who have undergone THA based on bearing materials. METHODS: In this prospective study, patients undergoing THA (181 patients, 187 hips) with metal-on-metal (MoM), hip resurfacing (HRA), metal-on-polyethylene (MoP), ceramic-on-polyethylene, ceramic-on-ceramic, or modular neck designs having revision surgery (between October 2013 and June 2017) underwent preoperative MRI. A single reader blinded to the bearing surface made an assessment of the synovial response (Gwet's AC1, 0.65-0.97); these data were compared with semiquantitative histology of tissue samples by a single reader (Gwet's AC1, 0.92) and semiquantitative wear, corrosion, and fretting analysis of retrieved components using Goldberg scoring (Gwet's AC1, 0.60-0.79). Direct noncontact measurements of implant wear were also made. Correlations and analyses of variance were used to assess associations between metrics and differences by implant type, respectively. RESULTS: Correlations were found between MRI synovial thickness with severity of fretting and corrosion damage of the female head-neck trunnion of femoral stems in modular designs (ρ = 0.26 [95% confidence interval {CI}, 0.12-0.39]; p = 0.015, n = 185) and ALTR grade and volumetric wear in MoM bearings (ρ = 0.93 [95% CI, 0.72-0.98]; p < 0.001, n = 10). MRI synovial thickness was highest in patients identified with aseptic lymphocyte-dominated vasculitis-associated lesions and diffuse tissue necrosis. On MRI, MoP hips demonstrated a distinct polymeric synovial response, whereas HRA, MoM, and modular hips more commonly demonstrated ALTR. Hips classified as having a polymeric synovial response on MRI had a greater number of particles present in tissue samples. CONCLUSIONS: In this study, we demonstrated that MRI of THA can distinguish synovial responses that reflect the bearing type of the implanted THA and correlate to direct measurements of implant wear, corrosion, and fretting and histologic assessment of wear particles in periprosthetic tissues. MRI provides a means of direct, noninvasive visualization of the host-generated synovial response. Patients presenting with painful arthroplasties may be evaluated for the cause of their discomfort, specifically highlighting any concerning synovial reactions that would warrant more prompt surgical intervention. Future studies would benefit from a prospective evaluation of different implants to assess the natural longitudinal history of arthroplasty complications, including the development and prevalence of ALTR across bearing constructs. LEVEL OF EVIDENCE: Level III, diagnostic study.