翻修全髋关节置换术治疗严重髋臼缺损时多孔钽假体组件的稳定性:放射立体分析研究
该研究用放射立体分析(RSA)测量55个用于翻修全髋关节置换术治疗严重髋臼缺损的多孔钽髋臼组件的迁移,平均随访4年。48个组件迁移小于预测后期松动的阈值(>1 mm),7个超过阈值且均无下方螺钉固定,其中6个用于骨盆不连续。2年时下方螺钉固定组绝对中位近端平移为|0.3| mm,无下方螺钉组为|0.4| mm(p=0.04)。
要点
- 55个多孔钽组件中48个迁移小于预测后期松动的>1 mm阈值,50个末次随访时未再翻修。
- 7个超过平移阈值,均无下方螺钉固定,其中6个用于骨盆不连续,5个因症状性松动再翻修。
- 2年时下方螺钉固定组绝对中位近端平移|0.3| mm,无螺钉组|0.4| mm(p=0.04)。
核心结论
- 经RSA测量,多数用于翻修全髋关节置换术治疗严重髋臼缺损的多孔钽髋臼组件早期迁移可接受,这预示这些组件长期生存率良好。
- 使用下方螺钉进一步改善了髋臼组件的固定。
研究设计与样本
- 采用放射立体分析(RSA)测量55个用于治疗严重髋臼缺损的多孔钽组件(28个Paprosky IIIA型,27个Paprosky IIIB型;21髋伴骨盆不连续)的迁移,平均随访4年(范围2至12年)。
- 证据等级:治疗性III级。
主要发现
- 55个组件中48个迁移小于预测后期松动的阈值(>1 mm),50个在末次随访时未再翻修。
- 7个组件超过平移阈值,均无下方螺钉固定,其中6个用于骨盆不连续,这6个中5个因患者症状相关松动而再翻修。
- 2年时,下方螺钉固定组绝对中位近端平移为|0.3| mm(范围|0.1|至|0.9| mm),无下方螺钉组为|0.4| mm(范围|0.03|至|16.4| mm)(p=0.04)。
关键数值与推荐
- 先前建立的2年内可接受近端平移阈值为≤1 mm。
- 预测后期松动的迁移阈值为>1 mm。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The acetabular components used in revision total hip arthroplasty (THA) to treat severe acetabular bone defects have high rates of re-revision at mid to long-term follow-up. Early translation of acetabular components used in revision THA is a good predictor of later loosening, and radiostereometric analysis (RSA) is the most sensitive method to measure migration. The objectives of the present study were to use RSA to compare the migration of the porous tantalum acetabular components used to treat severe bone defects with the previously established acceptable proximal translation threshold of ≤1 mm within 2 years, and to determine the effect on migration of the addition of inferior screws through the component into the ischium or pubis. METHODS: RSA was utilized to measure the migration of 55 porous tantalum components used to treat severe acetabular defects (28 Paprosky IIIA, 27 Paprosky IIIB; 21 hips with pelvic discontinuity) at a mean follow-up of 4 years (range, 2 to 12 years). RESULTS: Forty-eight of the 55 components migrated less than the threshold that predicts later loosening (>1 mm) and 50 had not been re-revised at the time of the latest follow-up. Seven components, none of which had inferior screw fixation, exceeded the translation threshold. Of these, 6 were implanted to treat pelvic discontinuity. Of those 6 components, 5 were re-revised for loosening related to patient symptoms. At 2 years, the absolute median proximal translation of components with inferior screw fixation was |0.3| mm (range, |0.1| to |0.9| mm), compared with |0.4| mm (range, |0.03| to |16.4| mm) for those without inferior screws (p = 0.04). CONCLUSIONS: As measured with use of RSA, the majority of porous tantalum acetabular components used in a revision THA to treat severe acetabular defects had acceptable early migration. This predicts good long-term survivorship of these components. The use of inferior screws further improved acetabular component fixation. LEVEL OF EVIDENCE: Therapeutic Level III. See Instructions for Authors for a complete description of levels of evidence.