EOS微剂量协议:人体模体中的首次全脊柱辐射剂量测量及与EOS标准剂量和常规数字放射摄影的比较

EOS Micro-dose Protocol: First Full-spine Radiation Dose Measurements in Anthropomorphic Phantoms and Comparisons with EOS Standard-dose and Conventional Digital Radiology.
PubMed 文献 · Spine (Phila Pa 1976) · 2018-01-01

本研究在代表5岁儿童和15岁青少年的两种人体模体中,首次报告EOS微剂量协议全脊柱检查的器官剂量和有效剂量,并与EOS标准剂量协议及常规数字放射摄影(CR)比较。青少年模体中微剂量较CR有效剂量降低17倍(94%),儿童模体中降低5倍(81%);但儿童模体中EOS标准剂量较CR未观察到有效剂量降低。

要点

  • 青少年模体中,微剂量协议有效剂量较CR降低17倍(94%)。
  • 儿童模体中,微剂量协议有效剂量较CR降低5倍(81%)。
  • 儿童模体中,EOS标准剂量协议较CR未观察到有效剂量降低。

核心结论

  • EOS成像微剂量选项使患者在全脊柱检查中接受的辐射剂量低于任何现有可用 modality。
  • 青少年模体中,与CR相比标准剂量协议的有效剂量降低符合预期;但儿童模体中,EOS标准剂量协议与参考CR系统相比未观察到有效剂量降低。

研究设计与样本

  • 研究为在人体模体中测量辐射剂量的比较研究。
  • 使用两种人体模体,分别代表5岁儿童和15岁青少年。
  • 模体在全脊柱成像中接受EOS微剂量和标准剂量协议曝光,并额外进行脊柱侧凸设置下的CR检查。
  • 所有模态均测量器官剂量,并使用热释光剂量计计算有效剂量。

主要发现

  • 青少年模体中,微剂量协议较CR系统有效剂量降低17倍(94%)。
  • 青少年模体中,微剂量较标准剂量协议有效剂量降低6倍(83%)。
  • 青少年模体中,标准剂量较CR系统有效剂量降低2.8倍(64%)。
  • 儿童模体中,微剂量协议较CR系统有效剂量降低5倍(81%)。
  • 儿童模体中,微剂量较标准剂量协议有效剂量降低7倍(86%);但EOS标准剂量协议较CR系统吸收剂量增加38%。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: A comparative study of radiation dose measured in anthropomorphic phantoms. OBJECTIVES: The aim of this study was to first report the first organ dose and effective dose measurements in anthropomorphic phantoms using the new EOS imaging micro-dose protocol in full-spine examinations, and to compare these measurements of radiation dose to measurements in the EOS standard-dose protocol and conventional digital radiology (CR). SUMMARY OF BACKGROUND DATA: Few studies evaluating organ dose and effective dose for the EOS low-dose scanner exist, and mainly for the standard-dose protocol. To the best of our knowledge, no studies of effective dose based on anthropomorphic phantom measurements exist for the new micro-dose protocol. METHODS: Two anthropomorphic phantoms, representing a 5-year-old (pediatric) and a 15-year-old (adolescent). The phantoms were exposed to EOS micro-dose and standard-dose protocols during full-spine imaging. Additionally, CR in scoliosis settings was performed. For all modalities, organ doses were measured and effective doses were calculated using thermoluminescent dosimeters. RESULTS: We found a 17-fold reduction (94%) of effective dose in micro-dose protocol compared with our CR system in the adolescent phantom. Micro-dose versus standard-dose protocol, showed a 6-fold reduction (83%), and for standard-dose versus our CR system a 2.8-fold reduction (64%) reduction of effective dose was observed.For the pediatric phantom, a 5-fold reduction (81%) of effective dose in micro-dose protocol compared to our CR system was observed. Micro-dose versus standard-dose protocol, showed a seven-fold (86%) reduction. However, we observed an increase in absorbed dose of 38% when comparing the EOS standard-dose protocol with our CR system. CONCLUSION: The EOS imaging micro-dose option exposes patients to lower radiation doses than any currently available modality for full-spine examination. Expected reduction of dose was established for the adolescent phantom when comparing CR and standard-dose protocol. However, no reduction of effective dose with EOS standard-dose protocol compared to our reference CR system was observed in the pediatric phantom. LEVEL OF EVIDENCE: N/A.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。

引用

  1. Pedersen PH, Petersen AG, Østgaard SE, et al. EOS Micro-dose Protocol: First Full-spine Radiation Dose Measurements in Anthropomorphic Phantoms and Comparisons with EOS Standard-dose and Conventional Digital Radiology. Spine (Phila Pa 1976). 2018. PMID: 30383725. doi:10.1097/BRS.0000000000002696