单膝置换设计中松质骨钛涂层非骨水泥与骨水泥胫骨部件的9年结局对比研究
两组9年累积成功率均为100%,无植入物相关失败,临床与影像学评分在各时间点均无差异。
要点
- 两组9年累积成功率均为100%,无植入物、术者或患者相关失败。
- 各时间点KSS、WOMAC、SF-12及Oxford膝关节评分两组间均无差异。
- 两组影像学评估参数均满意,非骨水泥松质骨钛涂层胫骨植入物无植入物相关失败。
核心结论
- 两种版本的该设计均获得满意的中期临床与影像学结局,非骨水泥松质骨钛涂层胫骨植入物未观察到植入物相关失败。
研究设计与样本
- 回顾性评估前瞻性收集的数据,A组54例(54膝,18男36女,手术时平均年龄63.2±5.2岁)植入非骨水泥部件,B组54例(54膝,17男37女,手术时平均年龄63.8±5.1岁)植入骨水泥部件,两组按年龄、性别、侧别、体重指数和随访时长匹配。
主要发现
- 两组所有患者均完成末次随访,平均随访8.6±0.4年。
- 9年生存分析显示两组在所有终点上累积成功率均为100%,两组均未观察到植入物相关、术者相关或患者相关失败。
- 所有时间间隔内,两组在Knee Society系统、Western Ontario and McMaster大学骨关节炎指数、Short Form-12和Oxford膝关节评分上均无差异记录。
- 影像学检查中两组所有评估参数均满意。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The cemented Advance Medial-Pivot total knee arthroplasty (TKA) was designed to reflect contemporary knee kinematics data and has shown satisfactory long-term outcomes. METHODS: We retrospectively evaluated prospectively collected data from 2 groups of patients. Group A consisted of 54 patients (54 TKAs), 18 men and 36 women, and mean age at surgery was 63.2 ± 5.2 years; group B consisted of 54 patients (54 TKAs), 17 men and 37 women, and mean age at surgery was 63.8 ± 5.1 years. Patients of both groups were matched for age, gender, side, body mass index, and length of follow-up. The cementless components of this design were implanted in group A and the cemented in group B. Implant failure, complication rates, clinical (both subjective and objective) and radiological outcomes were assessed in all patients of both groups. RESULTS: All patients of both groups were available for final follow-up evaluation at a mean of 8.6 ± 0.4 years. Survival analysis at 9 years showed a cumulative success rate of 100% in both groups with all end points. In neither group were implant-related, surgeon-related, or patient-related failures observed. When both groups were compared, in all time intervals, no differences were recorded on Knee Society system, Western Ontario and McMaster University Osteoarthritis Index, Short Form-12, and Oxford Knee Scores. On radiological examination, for both groups, all parameters evaluated were satisfactory. CONCLUSION: This study presents satisfactory midterm clinical and radiological outcomes with the use of both versions of this design. Moreover, no implant-related failures were observed with the use of cancellous titanium-coated tibial implants.
引用
- Karachalios T, Komnos G, Amprazis V, et al. A 9-Year Outcome Study Comparing Cancellous Titanium-Coated Cementless to Cemented Tibial Components of a Single Knee Arthroplasty Design. J Arthroplasty. 2018. PMID: 30077469. doi:10.1016/j.arth.2018.07.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。