第一代高交联聚乙烯在初次全髋关节置换术中的磨损分析:长达18年随访

Up to 18-Year Follow-Up Wear Analysis of a First-Generation Highly Cross-Linked Polyethylene in Primary Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2018-06-18

本研究对38例患者40髋初次全髋关节置换术后使用Crossfire高交联聚乙烯内衬的长期磨损进行随访,临床随访平均12.9年(范围7-18年),线性磨损率为0.056±0.036 mm/y,10年以上影像学随访患者中未见骨溶解,仅1例因机械性跌倒需翻修。结论认为长期磨损率与中期研究报道一致。

要点

  • 临床随访平均12.9年(范围7-18年),线性磨损率0.056±0.036 mm/y。
  • 影像学随访超过10年的患者中未观察到骨溶解。
  • 仅1例患者因机械性跌倒后需翻修手术。

核心结论

  • 本研究表明高交联聚乙烯内衬的长期磨损率继续与中期研究发表的磨损率相符。
  • 此前该队列10年平均年磨损率为0.05 mm/y,本次长达18年随访显示相似的磨损率。

研究设计与样本

  • 长期随访研究,纳入先前中期研究已报道的38例患者共40髋。
  • 患者均在1999年3月至2004年8月期间接受使用Crossfire高交联聚乙烯内衬的初次全髋关节置换术。
  • 计算该队列的年磨损率(mm/y),并联系患者询问自索引手术以来的并发症或翻修手术情况。

主要发现

  • 临床随访平均12.9年,范围7-18年;平均随访时长12.5年,范围10-17年。
  • 线性磨损率为0.056±0.036 mm/y。
  • 影像学随访超过10年的患者中未观察到骨溶解。
  • 仅1例患者在机械性跌倒后需要翻修手术。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The advent of highly cross-linked polyethylene (HCLPE) has significantly improved total hip arthroplasty survivorship. HCLPE has been shown to improve wear properties in midterm outcomes when compared to traditional polyethylene liners; however, there is a paucity of studies evaluating long-term outcomes. In addition, there is concern that wear rates may accelerate as the implant ages. Thus, the aims of this study are to report on the longest-to-date follow-up of a specific first-generation HCLPE liner and to determine whether there is a change in the annual wear rate over time. METHODS: Forty hips in 38 patients which were previously reported on in a midterm study were included in this long-term follow-up study. Patients in this cohort all received total hip arthroplasty between March 1999 and August 2004 using the Crossfire HCLPE liner. Annual wear rates (mm/y) were calculated for this cohort. Patients were contacted and asked about complications or revision procedures they may have had since the index procedure. RESULTS: Clinical follow-up averaged 12.9 years with a range of 7-18 years. The average follow-up duration was 12.5 years with a range of 10-17 years. Linear wear was found to be 0.056 ± 0.036 mm/y. Osteolysis was not observed in any of the patients with greater than 10-year radiographic follow-up. Furthermore, only 1 patient required revision surgery following a mechanical fall. CONCLUSION: Our study demonstrates the long-term wear rates associated with HCLPE liners continue to match rates published in midterm studies. Previously, we have reported that this cohort had an average annual wear rate of 0.05 mm/y over 10 years. This most recent report demonstrates a similar wear rate with up to 18-year follow-up.