青少年特发性脊柱侧凸患者全脊柱X线检查中铜滤过降低剂量方案

Dose Reduction Protocol for Full Spine X-ray Examination Using Copper Filters in Patients With Adolescent Idiopathic Scoliosis.
PubMed 文献 · Spine (Phila Pa 1976) · 2019-01-01

前瞻性病例系列研究评估青少年特发性脊柱侧凸(AIS)患者全脊柱X线检查使用铜滤过降低辐射剂量的新方案。体模研究比较无滤过与0.1、0.2、0.3毫米铜滤过的表面剂量;AIS术后随访患者交替使用含或不含0.2毫米铜滤过方案,由三名脊柱外科医生独立评估图像质量。使用0.2毫米铜滤过的全脊柱X线片可减少超过60%的辐射暴露,同时保持图像质量。

要点

  • 使用0.2毫米铜滤过可使AIS患者全脊柱X线检查辐射暴露减少超过60%。
  • 正位图像上3级评分比例无滤过为85%、滤过为75%,两者无显著差异。
  • 侧位图像上滤过组在Th2和Th12的3级评分频率显著低于无滤过组。

核心结论

  • AIS患者全脊柱X线检查使用0.2毫米铜滤过可减少超过60%的辐射暴露,同时保持图像质量。

研究设计与样本

  • 前瞻性病例系列。研究1:使用人体模体获取全脊柱X线片,比较无滤过与0.1、0.2、0.3毫米厚铜滤过的辐射剂量。研究2:对接受后路融合术的AIS患者随访时进行影像学评估,含或不含0.2毫米铜滤过的X线方案在连续随访间交替使用;由三名脊柱外科医生采用三级评分系统独立评估图像质量,正位视图评估6个点、侧位(左右)视图评估7个点。

主要发现

  • 研究1:无滤过正位和侧位表面剂量分别为0.31和0.93毫戈瑞;0.1毫米铜滤过分别为0.16和0.52毫戈瑞,0.2毫米铜滤过分别为0.11和0.36毫戈瑞,0.3毫米铜滤过分别为0.08和0.27毫戈瑞。
  • 研究2:AIS患者正位图像3级评分(两端终板均可辨认)比例无滤过为85%、滤过为75%,两种方案间无显著差异。
  • 研究2:侧位图像上,滤过组在Th2和Th12的3级评分频率显著低于无滤过组。

关键数值与推荐

  • 研究1中无滤过正位和侧位表面剂量分别为0.31和0.93毫戈瑞,0.1毫米铜滤过分别为0.16和0.52毫戈瑞,0.2毫米铜滤过分别为0.11和0.36毫戈瑞,0.3毫米铜滤过分别为0.08和0.27毫戈瑞。
  • 研究2正位图像3级评分比例无滤过为85%、滤过为75%,两者无显著差异。
  • 研究2侧位图像滤过组在Th2和Th12的3级评分频率显著低于无滤过组。

临床怎么用

  • AIS患者全脊柱X线检查采用0.2毫米铜滤过可减少超过60%的辐射暴露,同时保持图像质量。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: A prospective case series. OBJECTIVE: The aim of this study was to assess a new protocol for full spine X-ray using copper (Cu) filters to reduce radiation exposure in patients with adolescent idiopathic scoliosis (AIS). SUMMARY OF BACKGROUND DATA: Radiation exposure is associated with an increased risk of cancer development in children. To reduce the radiation exposure without compromising the image quality using existing radiographic equipment, a new computed radiography protocol was optimized using a variety of heavy metal filters. METHODS: Study 1: Whole spine radiographs were obtained using a human body phantom, and radiation doses without and with 0.1, 0.2, and 0.3 mm thick Cu filters were compared. Study 2: Patients with AIS who underwent posterior fusion were radiographically evaluated at follow-ups; the X-ray protocols with or without the use of 0.2-mm Cu filters were alternated between consecutive follow-ups. The image quality was independently evaluated using six points in the anterior-posterior (AP) view and seven in the lateral [left-right (LR)] view by three spine surgeons using a three-point grading system. RESULTS: Study 1: The surface doses while obtaining nonfiltered X-rays in AP and LR views were 0.31 and 0.93 mGy, respectively, whereas those with 0.1-, 0.2-, and 0.3-mm Cu filters were 0.16 and 0.52, 0.11 and 0.36, and 0.08 and 0.27 mGy, respectively.Study 2: In patients with AIS, the percentage of grade 3 scores (both endplates were identifiable) on AP-view images was 85% with nonfiltered X-rays and 75% with the filtered X-rays. However, there were no significant differences between the two protocols. On LR images, the frequency of grade 3 scores was significantly lower at Th2 and Th12 on filtered images than on nonfiltered ones. CONCLUSION: Whole spine radiographs using 0.2-mm Cu filters in patients with AIS could reduce radiation exposure more than 60% while preserving the image quality. LEVEL OF EVIDENCE: 4.