取出的磁控生长棒的力量测试
本研究对来自英国7个和丹麦1个中心的25例患者的45根取出的MAGEC生长棒进行实验室力量测试,并与2根未使用棒及制造商标准(42 Ibf)比较。结果显示64%的取出棒无力量输出,仅22%达到或超过制造商标准;体内留置时间与测试力量呈显著负相关(r=-0.63,P<0.005),植入超过38个月的12根棒均无力量输出。作者认为这些发现对植入物寿命提出疑问,并建议进一步开展临床结局研究。
要点
- 45根取出的MAGEC生长棒中,29根(64%)无力量输出,10根(22%)力量≥制造商标准,6根(13%)有部分力量但低于标准。
- 体内留置时间与测试力量呈显著负相关(r=-0.63,P<0.005);植入超过38个月的12根棒均无任何力量输出。
- 作者建议进一步开展临床结局研究,并对植入物寿命提出疑问。
核心结论
- 这是首项评估取出MAGEC生长棒力量(因而可能反映其功能)的研究;大多数取出棒无力量输出,其余力量降低,对植入物寿命提出疑问,并建议进一步开展临床结局研究。
研究设计与样本
- 对取出的MAGEC生长棒进行实验室分析,棒来自英国7个和丹麦1个中心,共45根棒来自25例病例,并与2根未使用棒及制造商定义标准(42 Ibf)比较。
主要发现
- 45根取出棒中,10根(22%)力量≥制造商标准,平均46.7(2.7)Ibf;6根(13%)有部分力量但低于标准,平均34.8(3.6)Ibf;29根(64%)无力量输出。
- 2根未使用MAGEC棒产生的平均力量分别为45.3(0.25)Ibf和50.2(1.4)Ibf,高于制造商标准。
- 体内留置时间与测试力量呈显著负相关(r=-0.63,P<0.005);植入超过38个月的12根棒均无任何力量输出。
- 植入时平均年龄8.6岁,棒在体内平均留置2.7年,以双棒构型为主。
关键数值与推荐
- 制造商定义的标准为42 Ibf。
- 力量≥制造商标准:10根(22%),平均46.7(2.7)Ibf。
- 有部分力量但低于制造商标准:6根(13%),平均34.8(3.6)Ibf。
- 无力量输出:29根(64%)。
- 未使用棒平均力量:45.3(0.25)Ibf和50.2(1.4)Ibf。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 证据等级为4级。
- 作者建议进一步开展临床结局研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
STUDY DESIGN: Laboratory analysis of explanted MAGnetic Expansion Control (MAGEC) growing rods. OBJECTIVE: The aim of this study was to measure the force produced by explanted MAGEC rods compared with new rods and assess the influence of clinical variables. SUMMARY OF BACKGROUND DATA: MAGEC rods are increasingly used in early-onset scoliosis. Some data are available describing the structure of explanted MAGEC rods, but to date, no study has assessed their function. METHODS: Explanted MAGEC rods were received from seven UK and one Danish center. The force produced by explanted rods on activation with the external remote controller was measured using a dedicated jig. Forces were compared with two unused rods as well as the manufacturer's defined standard (42 Ibf). Clinical variables were collected from contributing centers where possible and correlated with the force measurements. RESULTS: Forty-five MAGEC rods from 25 cases were received for analysis. The mean age at insertion was 8.6 years and rods were in vivo for a mean of 2.7 years in predominantly dual rod constructs. Two unused MAGEC rods produced a mean force of 45.3 (0.25) and 50.2 (1.4) Ibf, above the manufacturer's stated standard. Of the 45 explanted rods, 10 (22%) produced force greater or equal to manufacturer's standard, mean 46.7 (2.7) Ibf. Six rods (13%) produced some force but less than the manufacturer's standard, mean 34.8 (3.6) Ibf. Twenty-nine rods (64%) produced no force. The duration the rods were in vivo was significantly negatively correlated with the force produced on testing (r = -0.63, P < 0.005). Of the 12 rods implanted longer than 38 months, none produced any force. CONCLUSION: This is the first study of the force, and hence likely function, of explanted MAGEC rods. The majority of explanted rods produced no force, while others produced reduced force. These findings raise questions regarding the longevity of the implant and further clinical outcome studies are recommended. LEVEL OF EVIDENCE: 4.
引用
- Rushton PRP, Smith SL, Forbes L, et al. Force Testing of Explanted Magnetically Controlled Growing Rods. Spine (Phila Pa 1976). 2019. PMID: 30044365. doi:10.1097/BRS.0000000000002806
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。