脊柱手术中的椎弓根螺钉设计:是否存在差异?初次与翻修拔出强度的生物力学研究

Pedicle Screw Designs in Spinal Surgery: Is There a Difference? A Biomechanical Study on Primary and Revision Pull-Out Strength.
PubMed 文献 · Spine (Phila Pa 1976) · 2019-01-01

该研究在标准化聚氨酯泡沫模型中测试单螺纹、混合螺纹和双螺纹椎弓根螺钉的初次及翻修拔出强度。结果显示双螺纹螺钉初次拔出强度最高(2628.8 N),混合螺纹螺钉翻修拔出强度显著降低18.6%(1890.2 N,P=0.0173)。作者认为混合螺纹螺钉需翻修时可考虑使用更大直径螺钉。

要点

  • 双螺纹螺钉初次拔出强度(2628.8 N)高于单螺纹螺钉(2184.4 N,P<0.05)。
  • 混合螺纹螺钉翻修拔出强度降低18.6%(1890.2 N,P=0.0173),单螺纹和双螺纹螺钉无显著降低。
  • 使用更大直径螺钉翻修可使拔出强度恢复至基线水平。

核心结论

  • 双螺纹螺钉提供最强的初次拔出强度。
  • 混合螺纹螺钉翻修拔出强度最差,较其他螺钉设计降低18.6%。
  • 作者认为,混合螺纹螺钉需翻修时(初次或翻修手术),术者可考虑使用更大直径螺钉以提高拔出强度。

研究设计与样本

  • 实验性实验室生物力学研究,在标准化聚氨酯泡沫中测试三种椎弓根螺钉设计的拔出强度。
  • 测试三种序列:序列1为6.5 mm螺钉置入后测初次拔出强度;序列2为6.5 mm螺钉置入、取出后重新置入测翻修拔出强度;序列3为6.5 mm螺钉置入、取出后改用同螺纹设计的7.5 mm螺钉重新置入测翻修拔出强度。
  • 证据等级为5级。

主要发现

  • 初次拔出强度:双螺纹螺钉2628.8 N vs 单螺纹螺钉2184.4 N(P<0.05)。
  • 翻修拔出强度:混合螺纹螺钉降低18.6%(1890.2 N,P=0.0173)。
  • 单螺纹和双螺纹螺钉翻修拔出强度无显著降低。
  • 使用更大直径螺钉翻修可使拔出强度恢复至基线水平。

临床怎么用

  • 作者认为,混合螺纹螺钉需翻修时(初次或翻修手术),术者可考虑使用更大直径螺钉以提高拔出强度。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: An experimental laboratory-based biomechanical study. OBJECTIVE: The objective of this study was to evaluate, in a synthetic bone model, the difference in primary and revision pull-out strength using pedicle screws of different thread designs. SUMMARY OF BACKGROUND DATA: Over the past few decades, there has been a growing interest in optimizing the screw pull-out strength using various screw designs (single-thread, mixed-thread, dual-thread). Although primary pull-out strength has been studied previously, little is known about revision pull-out strength of different pedicle screws. METHODS: The pull-out strengths of three different pedicle screw designs (single-thread, mixed-thread, dual-thread) were tested in standardized polyurethane foam in three sequences. Sequence 1: A 6.5 mm screw was inserted into the foam block and the primary pull-out strength measured. Sequence 2: A 6.5 mm screw was inserted, removed, and then reinserted into the same foam block. The revision pull-out strength was then measured. Sequence 3: A 6.5 mm screw was inserted, removed and a 7.5-mm screw of the same thread design was reinserted. The revision pull-out strength was then measured. RESULTS: The primary pull-out strength was similar across screw designs, although dual-thread screws showed higher primary pull-out strength (2628.8 N) compared to single-thread screws (2184.4 N, P < 0.05). For revision pull-out strength, the mixed-thread screws had significant reduction in revision pull-out strength of 18.6% (1890.2 N, P = 0.0173). Revision with a larger diameter screw improved the pull-out strength back to baseline. Single and dual-thread screws showed no significant reduction in revision pull-out strength. CONCLUSION: The dual-threaded screws provided the strongest primary pull-out strength for spinal fixation. The mixed-thread screws, however, had the poorest revision pull-out strength, decreasing by 18.6% compared to other screw designs. In cases in which mixed-threaded screws have to be revised (at the index or revision surgery), surgeons should consider the use of larger diameter screws to improve the pull-out strength. LEVEL OF EVIDENCE: 5.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。