当代全膝关节置换术股骨部件承重面钴铬损伤模式的发生率

What Is the Incidence of Cobalt-Chromium Damage Modes on the Bearing Surface of Contemporary Femoral Component Designs for Total Knee Arthroplasty?
PubMed 文献 · J Arthroplasty · 2018-05-17

通过多机构骨科植入物回收项目收集256例取出的钴铬股骨部件,评估第三体损伤和微尺度电腐蚀损伤。52%(134/256)存在轻至重度损伤,82%(209/256)疑似微尺度电腐蚀损伤。前后径比、松动和患者体重与损伤评分相关。

要点

  • 52%(134/256)的取出钴铬股骨部件存在轻至重度第三体损伤。
  • 82%(209/256)的部件疑似微尺度电腐蚀损伤。
  • 前后径比(β=-8.07;P<.001)、松动(β=-0.52;P=.006)和患者体重(β=0.01;P=.007)与损伤评分相关。

核心结论

  • 回收研究结果支持当代与长期全膝关节置换装置存在相似的损伤机制。
  • 观察到松动、前后径贴合度和患者体重与表面损伤增加相关。

研究设计与样本

  • 通过多机构骨科植入物回收项目收集256例取出的钴铬股骨部件,植入时间1-15年。
  • 部件主要因松动(84/256)、不稳定(62/256)和感染(46/256)而翻修。
  • 采用半定量评分法评估第三体损伤,用数字光学显微镜评估微尺度电腐蚀损伤(MECD)。
  • 测量代表性部件曲率半径以计算前后径比和内外径比,用白光干涉仪测量表面粗糙度。
  • 使用多变量线性模型检验损伤评分与植入物及患者因素的关联,用Spearman ρ相关检验确定粗糙度测量值与损伤评分的关联。

主要发现

  • 52%(134/256)的部件存在轻至重度损伤。
  • 82%(209/256)的部件疑似微尺度电腐蚀损伤。
  • 多变量线性模型中,前后径比(β=-8.07;P<.001)、松动(β=-0.52;P=.006)和患者体重(β=0.01;P=.007)与损伤评分相关。
  • Ra值(ρ=0.196;P=.002)和Rq值(ρ=0.157;P=.012)随损伤评分增加而增大。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The purpose of this study was to determine the incidence of metal release in contemporary total knee arthroplasty and the patient-related factors associated with this release. METHODS: In total, 256 retrieved cobalt-chromium femoral components were collected through a multi-institutional orthopedic implant retrieval program (implanted: 1-15 years). Implants were mainly revised for loosening (84/256), instability (62/256), and infection (46/256). Third-body damage was assessed using a semiquantitative scoring method. Microscale electro-corrosion damage (MECD) was evaluated using digital optical microscopy. Radii of curvature were measured from representative components to calculate anterior-posterior and medial-lateral ratios. Femoral component surface roughness was measured using a white light interferometer. Using a multivariable linear model, associations between damage score, implant, and patient factors were tested. Spearman's ρ correlation tests were performed to determine the association between roughness measurements and damage score. RESULTS: Mild to severe damage was observed in 52% (134/256) of the components. In the multivariable linear model, anterior-posterior ratio (β = -8.07; P < .001), loosening (β = -0.52; P = .006), and patient weight (β = 0.01; P = .007) were associated with damage score. Suspected MECD damage was observed in 82% (209/256) of components. The Ra value (ρ = 0.196; P = .002) and Rq value (ρ = 0.157; P = .012) increased as the damage score increased. CONCLUSION: The findings of this retrieval study support that similar damage mechanisms exist in contemporary and long-term total knee arthroplasty devices. Additionally, we observed associations between loosening, anterior-posterior conformity, and patient weight with increased surface damage.

来源:J Arthroplasty,查看原文。版权归原作者所有。