文献速递关节外科

固定平台与活动平台全膝关节置换术的随机对照试验:平均十年随访

A randomized controlled trial of fixed- versus mobile-bearing total knee arthroplasty: a follow-up at a mean of ten years.
PubMed 文献 · Bone Joint J · 2018-01-01

这项随机对照试验纳入240例接受骨水泥固定初次全膝关节置换术的患者,比较三种胫骨部件(全聚乙烯固定平台、组配式金属背衬固定平台和活动平台)在中位十年随访时的耐久性、活动度和功能。结果显示活动平台设计可靠耐久,但在最大屈膝度、功能和耐久性方面并未优于后稳定型固定平台设计。

要点

  • 活动平台与固定平台全膝关节置换术后十年因任何原因翻修的生存率无差异。
  • 十年时平均最大活动度无差异(p=0.8),膝关节学会功能评分无差异(p=0.63)。
  • 髌骨倾斜发生率无差异(p=0.12)。

核心结论

  • 在这项临床随机对照试验中,活动平台全膝关节置换设计可靠且耐久,但在术后十年并未提供更好的最大屈膝度、功能或耐久性。

研究设计与样本

  • 随机对照试验,共纳入240例接受骨水泥固定初次全膝关节置换术的患者,随机使用三种胫骨部件之一:全聚乙烯固定平台、组配式金属背衬固定平台和活动平台。
  • 患者在中位随访十年(四分位距9.2至10.4年)时接受评估。

主要发现

  • 以因任何原因翻修的生存率衡量的耐久性无差异。
  • 十年时平均最大活动度无差异(p=0.8)。
  • 以膝关节学会功能评分衡量的功能无差异(p=0.63)。
  • 髌骨倾斜发生率无差异(p=0.12)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: It has been suggested that mobile-bearing total knee arthroplasty (TKA) might lead to better outcomes by accommodating some femorotibial rotational mismatch, thereby reducing contact stresses and polyethylene wear. The aim of this study was to determine whether there is a difference between fixed- and mobile-bearing versions of a contemporary TKA with respect to durability, range of movement (ROM) and function, ten years postoperatively. PATIENTS AND METHODS: A total of 240 patients who were enrolled in this randomized controlled trial (RCT) underwent a primary cemented TKA with one of three tibial components (all-polyethylene fixed-bearing, modular metal-backed fixed-bearing and mobile-bearing). Patients were reviewed at a median follow-up of ten years (IQR 9.2 to 10.4). RESULTS: There was no difference in durability, as measured by survivorship free of revision for any reason, nor in mean maximal ROM at ten years (p = 0.8). There was also no difference in function, as measured by Knee Society (KS) function scores (p = 0.63) or the prevalence of patellar tilt (p = 0.12). CONCLUSION: In this clinical RCT, the mobile-bearing design of TKA was found to be reliable and durable, but did not provide better maximum knee flexion, function or durability ten years postoperatively compared with a posterior-stabilized, fixed-bearing design incorporating either an all-polyethylene or a modular-metal-backed tibial component. Cite this article: Bone Joint J 2018;100-B:925-9.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Abdel MP, Tibbo ME, Stuart MJ, et al. A randomized controlled trial of fixed- versus mobile-bearing total knee arthroplasty: a follow-up at a mean of ten years. Bone Joint J. 2018. PMID: 29954199. doi:10.1302/0301-620X.100B7.BJJ-2017-1473.R1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。