高交联聚乙烯磨损率与髋臼假体方位:至少十年随访
本研究回顾94例患者100髋初次全髋关节置换术后至少十年,采用放射立体分析测量髋臼假体前倾角、外倾角及聚乙烯磨损率。结果显示髋臼假体位于或超出安全区时磨损率无差异,高交联聚乙烯在关节面磨损方面似乎是一种宽容的承重材料,但仍需注意假体位置以确保稳定性。
要点
- 平均磨损率为0.055 mm/年(标准差0.060),恰好一半髋关节位于安全区内。
- 安全区内与安全区外髋臼假体的磨损率无差异(中位差值0.012 mm/年;p=0.091)。
- 仅达到目标前倾角、仅达到目标外倾角或两者均未达到时,磨损率与安全区内相比均无差异。
核心结论
- 将髋臼假体置于安全区内或安全区外,在长期随访中未改变高交联聚乙烯的磨损率至有骨溶解风险的水平。
- 高交联聚乙烯在关节面磨损方面似乎是一种宽容的承重材料,但仍须注意正确放置髋臼假体以使植入物稳定。
研究设计与样本
- 回顾性研究,纳入94例患者100髋,均为至少十年前接受初次全髋关节置换术者。
- 所有患者使用同一类型髋臼假体,承重偶件为28 mm钴铬合金头对高交联聚乙烯内衬。
- 采用仰卧位放射立体分析检查,获取前后位和侧位配对图像。
- 测量髋臼假体前倾角和外倾角,以及股骨头总穿透量,除以植入时间确定聚乙烯磨损率。
主要发现
- 平均前倾角为19.4°(范围-15.2°至48°,标准差11.4°)。
- 平均外倾角为43.4°(范围27.3°至60.5°,标准差6.6°)。
- 平均磨损率为0.055 mm/年(标准差0.060)。
- 恰好一半的髋关节位于安全区内。
- 安全区内与安全区外髋臼假体的磨损率无差异(中位差值0.012 mm/年;p=0.091)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: The aim of this study was to determine whether there is a difference in the rate of wear between acetabular components positioned within and outside the 'safe zones' of anteversion and inclination angle. PATIENTS AND METHODS: We reviewed 100 hips in 94 patients who had undergone primary total hip arthroplasty (THA) at least ten years previously. Patients all had the same type of acetabular component with a bearing couple which consisted of a 28 mm cobalt-chromium head on a highly crosslinked polyethylene (HXLPE) liner. A supine radiostereometric analysis (RSA) examination was carried out which acquired anteroposterior (AP) and lateral paired images. Acetabular component anteversion and inclination angles were measured as well as total femoral head penetration, which was divided by the length of implantation to determine the rate of polyethylene wear. RESULTS: The mean anteversion angle was 19.4° (-15.2° to 48°, sd 11.4°), the mean inclination angle 43.4° (27.3° to 60.5°, sd 6.6°), and the mean wear rate 0.055 mm/year (sd 0.060). Exactly half of the hips were positioned inside the 'safe zone'. There was no difference (median difference, 0.012 mm/year; p = 0.091) in the rate of wear between acetabular components located within or outside the 'safe zone'. When compared to acetabular components located inside the 'safe zone', the wear rate was no different for acetabular components that only achieved the target anteversion angle (median difference, 0.012 mm/year; p = 0.138), target inclination angle (median difference, 0.013 mm/year; p = 0.354), or neither target (median difference, 0.012 mm/year; p = 0.322). CONCLUSION: Placing the acetabular component within or outside the 'safe zone' did not alter the wear rate of HXLPE at long-term follow-up to a level that risked osteolysis. HXLPE appears to be a forgiving bearing material in terms of articular surface wear, but care must still be taken to position the acetabular component correctly so that the implant is stable. Cite this article: Bone Joint J 2018;100-B:891-7.