文献速递关节外科

原生与置换后膝关节的松弛度特征——机器人辅助间隙平衡全膝关节置换术的应用

Laxity Profiles in the Native and Replaced Knee-Application to Robotic-Assisted Gap-Balancing Total Knee Arthroplasty.
PubMed 文献 · J Arthroplasty · 2018-05-14

研究在14具尸体标本上实施机器人辅助间隙平衡全膝关节置换术,测量原生与置换后膝关节全程屈曲范围内的松弛度,并评估主动韧带张力技术实现目标间隙轮廓的精度。结果显示原生间隙自完全伸直至屈曲20°期间内侧增加3.4±1.7 mm、外侧增加3.7±2.1 mm(P<.001),此后保持恒定;置换后0°与90°屈曲间隙相等,但中屈曲期松弛度较0°和90°大2-4 mm(P<.001);计划与实际间隙的均方根误差内侧1.6 mm、外侧1.7 mm。

要点

  • 原生膝关节间隙自完全伸直至屈曲20°内侧增加3.4±1.7 mm、外侧增加3.7±2.1 mm(P<.001)。
  • 置换后0°与90°屈曲间隙相等,但中屈曲期松弛度较0°和90°大2-4 mm(P<.001)。
  • 机器人辅助控制韧带张力下,计划与实际间隙均方根误差内侧1.6 mm、外侧1.7 mm。

核心结论

  • 以0°和90°屈曲间隙相等为目标,可获得伸直与屈曲间隙相等但中屈曲期间隙更大的结果。
  • 在机器人辅助全膝关节置换术中采用控制性韧带张力,可实现与计划间隙轮廓一致的软组织平衡。

研究设计与样本

  • 在14具尸体标本上实施机器人辅助间隙平衡全膝关节置换术。
  • 切除胫骨近端后,用机器人张力器在全程屈曲范围内动态张紧软组织包膜测量原生胫股间隙。
  • 股骨假体对线以0°和90°屈曲间隙平衡为准,最终试模时用机器人张力器测量术后间隙并与计划间隙比较。

主要发现

  • 原生间隙自完全伸直至屈曲20°期间内侧增加3.4±1.7 mm、外侧增加3.7±2.1 mm(P<.001),此后至剩余屈曲范围保持恒定。
  • 置换后0°与90°屈曲间隙相等,但中屈曲期间隙松弛度较0°和90°大2-4 mm(P<.001)。
  • 计划间隙与实际测量术后间隙的均方根误差在全程活动范围内内侧为1.6 mm、外侧为1.7 mm。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The traditional goal of the gap-balancing method in total knee arthroplasty is to create equal and symmetric knee laxity throughout the arc of flexion. The purpose of this study was to (1) quantify the laxity in the native and the replaced knee throughout the range of flexion in gap-balancing total knee arthroplasty (TKA) and (2) quantify the precision in achieving a targeted gap profile throughout flexion using a robotic-assisted technique with active ligament tensioning. METHODS: Robotic-assisted, gap-balancing TKA was performed in 14 cadaver specimens. The proximal tibia was resected, and the native tibiofemoral gaps were measured using a robotic tensioner that dynamically tensioned the soft-tissue envelope throughout the arc of flexion. The femoral implant was then aligned to balance the gaps at 0° and 90° of flexion. The postoperative gaps were then measured during final trialing with the robotic tensioner and compared with the planned gaps. RESULTS: The native gaps increased by 3.4 ± 1.7 mm medially and 3.7 ± 2.1 mm laterally from full extension to 20° of flexion (P < .001) and then remained consistent through the remaining arc of flexion. Gap balancing after TKA produced equal gaps at 0° and 90° of flexion, but the gap laxity in midflexion was 2-4 mm greater than at 0° and 90° (P < .001). The root mean square error between the planned gaps and actual measured postoperative gaps was 1.6 mm medially and 1.7 mm laterally throughout the range of motion. CONCLUSION: Aiming for equal gaps at 0° and 90° of flexion produced equal gaps in extension and flexion with larger gaps in midflexion. Consistent soft-tissue balance to a planned gap profile could be achieved by using controlled ligament tensioning in robotic-assisted TKA.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Shalhoub S, Moschetti WE, Dabuzhsky L, et al. Laxity Profiles in the Native and Replaced Knee-Application to Robotic-Assisted Gap-Balancing Total Knee Arthroplasty. J Arthroplasty. 2018. PMID: 29909956. doi:10.1016/j.arth.2018.05.012

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。