人尸体膝关节模型中纤维增强半月板替代装置的挤压螺钉与缝线纽扣固定对比

Interference Screw Versus Suture Endobutton Fixation of a Fiber-Reinforced Meniscus Replacement Device in a Human Cadaveric Knee Model.
PubMed 文献 · Am J Sports Med · 2018-05-30

在7例人尸体膝关节(年龄17-61岁)中,于1个前部和2个潜在后部隧道位置,分别测试缝线纽扣与挤压螺钉固定新型半月板装置。结果显示缝线固定循环加载伸长更大,刚度仅为挤压螺钉固定的28%-37%,后者更接近原生半月板根部。

要点

  • 缝线固定循环加载伸长较挤压螺钉固定增加54.1%-150.7%,差异有统计学意义(P<.05)。
  • 缝线固定组刚度为36.5-41.6 N/mm,仅为挤压螺钉固定组(98.7-131.6 N/mm)的28%-37%。
  • 两种固定技术的极限拉伸载荷均与原生半月板根部处于相同范围。

核心结论

  • 挤压螺钉固定在伸长和刚度方面优于缝线固定,该发现应考虑用于新型半月板替代装置的设计和植入。

研究设计与样本

  • 控制性实验室研究,在7例人尸体膝关节(年龄17-61岁)中,于1个前部和2个潜在后部隧道位置分别测试缝线纽扣与挤压螺钉固定。

主要发现

  • 全部3个隧道中,缝线固定循环加载期间伸长较挤压螺钉固定增加54.1%-150.7%(P<.05),挤压螺钉固定接近原生根部。
  • 两种固定技术的极限拉伸载荷均与原生根部处于相同范围。
  • 缝线固定组刚度为36.5-41.6 N/mm,仅为挤压螺钉固定组(98.7-131.6 N/mm)的28%-37%,后者接近原生根部刚度(前部:175.4 ± 24.2 N/mm;后部:157.6 ± 22.9 N/mm)。

临床怎么用

  • 作者认为,挤压螺钉固定在伸长和刚度方面优于缝线固定,该发现应考虑用于新型半月板替代装置的设计和植入。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Meniscal lesions represent one of the most common intra-articular knee injuries. Meniscus replacement devices are needed to restore load distribution and knee stability after meniscectomy. Fixation of these devices is crucial to the generation of hoop stresses and the distribution of loads in the joint. PURPOSE: To evaluate 2 different fixation techniques (suture endobutton and interference screw) for implantation of a novel meniscus device. STUDY DESIGN: Controlled laboratory study. METHODS: In 7 human cadaveric knees (aged 17-61 years), 1 anterior and 2 potential posterior tunnel locations were investigated, and both fixation techniques were tested in each tunnel. The native meniscus roots, devices fixed with a suture endobutton, and devices fixed with an interference screw were gripped with cryoclamps, and tibias were drilled and loaded into a custom jig. Samples were preloaded, preconditioned, loaded for 500 cycles (50-150 N), and tested in tension until failure. RESULTS: For all 3 tunnels, suture fixation resulted in greater elongation (54.1%-150.7% greater; P < .05) during cyclic loading than interference screw fixation, which approximated the native roots. Both fixation techniques displayed ultimate tensile loads in the same range as native roots. However, stiffness of the suture fixation groups (36.5-41.6 N/mm) was only 28% to 37% of that of the interference screw fixation groups (98.7-131.6 N/mm), which had values approaching those of the native roots (anterior: 175.4 ± 24.2 N/mm; posterior: 157.6 ± 22.9 N/mm). CONCLUSION: Interference screw fixation was found to be superior to suture fixation with regard to elongation and stiffness, a finding that should be considered in the design and implantation of novel meniscus replacement devices. CLINICAL RELEVANCE: With the emergence of various devices for total meniscus replacement, the establishment of fixation strategies is crucial for the generation of tensile hoop stresses and the efficacy of these approaches.