可调节环股骨皮质悬吊固定用于髌腱前交叉韧带重建:与界面螺钉固定的时间零点生物力学比较

Adjustable-Loop Femoral Cortical Suspensory Fixation for Patellar Tendon Anterior Cruciate Ligament Reconstruction: A Time Zero Biomechanical Comparison With Interference Screw Fixation.
PubMed 文献 · Am J Sports Med · 2018-05-15

研究比较可调节环皮质纽扣与金属界面螺钉在猪股骨远端固定人骨-髌腱-骨异体移植物的时间零点生物力学特性,每组8例。两组失败载荷与线性刚度无显著差异,但可调节环皮质纽扣组循环位移更高(2.1±0.6 mm vs 1.3±0.4 mm,P=0.01)。作者认为可调节环皮质纽扣或可作为前交叉韧带重建中骨-髌腱-骨移植物股骨固定的可接受替代方案。

要点

  • 可调节环皮质纽扣与界面螺钉的失败载荷(700±256 N vs 688±215 N,P=0.92)和线性刚度(219±48 N/mm vs 218±49 N/mm,P=0.97)无显著差异。
  • 可调节环皮质纽扣组循环位移高于界面螺钉组(2.1±0.6 mm vs 1.3±0.4 mm,P=0.01)。
  • 失败模式不同:界面螺钉组常见骨块滑移(n=5),可调节环皮质纽扣组常见骨块经缝线孔骨折(n=6)。

核心结论

  • 可调节环皮质纽扣与界面螺钉的时间零点失败载荷相似,但可调节环皮质纽扣的循环位移更高,位移平均差异小于1 mm。
  • 作者认为可调节环皮质纽扣或可作为前交叉韧带重建中骨-髌腱-骨移植物股骨固定的可接受替代方案,但循环位移差异的临床意义未知,应在未来研究中评估。

研究设计与样本

  • 控制实验室研究,可调节环皮质纽扣组n=8,金属界面螺钉组n=8,使用配对的人骨-髌腱-骨异体移植物固定于猪股骨远端。
  • 构建物先进行预适应(10 N至50 N,1 Hz,10个循环),再循环加载(50 N至250 N,1 Hz,500个循环),最后以20 mm/min拉伸至失败。

主要发现

  • 可调节环皮质纽扣组与金属界面螺钉组的失败载荷分别为700±256 N与688±215 N,P=0.92,无显著差异。
  • 可调节环皮质纽扣组与金属界面螺钉组的线性刚度分别为219±48 N/mm与218±49 N/mm,P=0.97,无显著差异。
  • 可调节环皮质纽扣组循环位移为2.1±0.6 mm,高于界面螺钉组的1.3±0.4 mm,P=0.01。
  • 界面螺钉组最常见的失败机制为骨块滑移(n=5),可调节环皮质纽扣组最常见的失败机制为骨块经缝线孔骨折(n=6)。

临床怎么用

  • 作者认为可调节环皮质纽扣或可作为前交叉韧带重建中骨-髌腱-骨移植物股骨固定的可接受替代方案。
  • 作者指出观察到的循环位移差异的临床意义未知,应在未来研究中评估。

局限性

  • 作者指出循环位移差异的临床意义未知,应在未来研究中评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Adjustable-loop cortical buttons for femoral fixation of bone-tendon-bone grafts have potential advantages over interference screw fixation; however, these devices have not been benchmarked biomechanically against interference screws. Purpose/Hypothesis: The purpose was to compare the time zero biomechanical properties of commercially available, adjustable-loop cortical button and metallic interference screws for femoral fixation of bone-tendon-bone grafts. It was hypothesized that no significant differences would be found in biomechanical properties between fixation techniques. STUDY DESIGN: Controlled laboratory study. METHODS: Adjustable-loop cortical buttons (n = 8) and metallic interference screws (n = 8) were used to fix matched pairs of human bone-tendon-bone allografts in porcine distal femurs. These constructs were preconditioned (10 N to 50 N at 1 Hz, 10 cycles), subjected to cyclic loading (50 N to 250 N at 1 Hz, 500 cycles), and then pulled to failure at 20 mm/min. RESULTS: The loads to failure (mean ± SD, 700 ± 256 N vs 688 ± 215 N, P = .92) and linear stiffnesses (219 ± 48 N/mm vs 218 ± 49 N/mm, P = .97) for the adjustable-loop cortical button and metallic interference screws, respectively, were not significantly different. Cyclic displacement was higher in the adjustable-loop cortical button group (2.1 ± 0.6 mm vs 1.3 ± 0.4 mm, P = .01). The mechanism of failure was different between groups, with bone block slippage occurring most commonly in the interference screw group (n = 5) and fracture of the bone block through the suture hole occurring most commonly in the adjustable-loop cortical button group (n = 6). CONCLUSION: Adjustable-loop cortical buttons and interference screws have similar time zero failure loads, although cyclic displacement was higher with the adjustable-loop cortical buttons. The mean difference in displacement was less than 1 mm compared with the interference screw. CLINICAL RELEVANCE: Adjustable-loop cortical buttons may be an acceptable alternative to an interference screw for femoral fixation of bone-tendon-bone grafts in anterior cruciate ligament reconstruction. The clinical relevance of the observed differences in cyclic displacement is unknown and should be evaluated in future studies.