计算机断层扫描技术有助于理解取出的全膝关节置换假体的磨损模式

Computed Tomography Techniques Help Understand Wear Patterns in Retrieved Total Knee Arthroplasty.
PubMed 文献 · J Arthroplasty · 2018-04-16

本研究分析30例取出的现代全膝关节置换假体,术前基于三维CT图像测量假体在冠状面的位置,并对取出的聚乙烯部件行微型CT扫描重建三维模型,计算内外侧厚度差。结果显示胫股角与股骨角均与聚乙烯间室厚度差显著相关:内翻与内侧间室更薄强相关,外翻与外侧间室更薄相关;欠佳的胫股取向与胫骨部件角度与更大的聚乙烯间室厚度差相关。

要点

  • 内翻角与内侧聚乙烯间室更薄强相关,外翻角与外侧间室更薄相关。
  • 欠佳的胫股取向和胫骨部件角度与聚乙烯间室厚度差增大相关。
  • 研究采用基于微型CT的取出物分析结合术前三维CT测量假体位置。

核心结论

  • 组件位置影响聚乙烯损伤,有助于理解这些假体的体内性能。

研究设计与样本

  • 分析30例取出的现代全膝关节置换假体,术前基于三维CT图像用诊断算法测量假体在冠状面的位置。
  • 对每件取出的聚乙烯部件行微型CT扫描并数字重建高分辨率计算三维模型,计算内外侧厚度差。
  • 进行统计分析以探究组件位置与损伤模式之间的关联。

主要发现

  • 胫股角与股骨角均与聚乙烯间室厚度差显著相关:内翻角与内侧间室更薄强相关,外翻角与外侧间室更薄相关。
  • 欠佳的胫股取向与胫骨部件角度与聚乙烯间室厚度差增大相关。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Suboptimal total knee arthroplasty (TKA) position of both femoral and tibial components is thought to be linked with poor clinical outcomes, polyethylene wear and the "unexplained" painful knee arthroplasty. The aim of this study was to better understand the effect of implant orientation on knee implant performance. METHODS: We analyzed 30 retrieved contemporary TKA implants. Implant positioning measurements in the coronal plane were made prior to revision using a diagnostic algorithm, based on 3D computed tomography (CT) images. Each retrieved polyethylene component was imaged using a micro-CT scanner and a high resolution computational 3D model of each component was digitally reconstructed. The difference in thickness between medial and lateral components was calculated. Statistical analysis was performed to investigate the association between component positioning and damage patterns. RESULTS: We found a significant correlation between both the tibiofemoral and femoral angles and difference in thickness between polyethylene compartments: varus angulations were strongly associated with thinner medial compartments, whilst valgus angulations were associated with thinner lateral compartments. Moreover, suboptimal tibiofemoral orientations and tibial component angulations were associated to greater differences in thickness between polyethylene compartments. CONCLUSION: Our study is the first to compare accurate 3D CT measurements of prerevision TKA positioning in the coronal plane with postrevision retrieval analysis from innovative, accurate and highly reliable micro-CT-based method. Our results demonstrate the impact of component positioning on polyethylene damage and helps understanding of the in vivo performance of these implants. LEVEL OF EVIDENCE: III.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Cerquiglini A, Henckel J, Hothi HS, et al. Computed Tomography Techniques Help Understand Wear Patterns in Retrieved Total Knee Arthroplasty. J Arthroplasty. 2018. PMID: 29754982. doi:10.1016/j.arth.2018.04.010