传感器辅助软组织测量在初次全膝关节置换术中的可靠性
本研究在33例初次全膝关节置换术(TKA)常规技术平衡后,用无线传感器连续3次测量股胫压力(2次盲法、1次可视反馈),评估该设备的可靠性。结果显示盲法测量间组内相关系数12项中2项差、4项中等、6项良好;盲法与可视测量间2项差、6项中等、4项良好;10°屈曲位测量可靠性较差。结论为该无线传感器在83%的测量中具有中等至良好可靠性。
要点
- 无线传感器在83%的测量中具有中等至良好可靠性。
- 盲法测量间12项中2项差、4项中等、6项良好。
- 10°屈曲位测量的可靠性较差。
核心结论
- 无线传感器在83%的测量中具有中等至良好可靠性。
- 10°屈曲位测量与较差的可靠性相关。
研究设计与样本
- 研究纳入33例初次全膝关节置换术(TKA),常规技术平衡后使用无线传感器测量股胫压力。
- 每例TKA在膝关节10°、45°、90°屈曲位测量内外侧间室压力,共6次测量。
- 另计算3个位置上的合并压力及间室间压力差,每例TKA共另6次测量。
- 盲法测量进行2次(n=29),可视反馈测量进行1次(n=32)。
主要发现
- 盲法测量间组内相关系数:12项中2项(17%)差、4项(33%)中等、6项(50%)良好。
- 盲法与可视测量间组内相关系数:12项中2项(17%)差、6项(50%)中等、4项(33%)良好。
- 10°屈曲位测量与较差的可靠性相关。
关键数值与推荐
- 盲法测量间组内相关系数:差2/12(17%)、中等4/12(33%)、良好6/12(50%)。
- 盲法与可视测量间组内相关系数:差2/12(17%)、中等6/12(50%)、良好4/12(33%)。
- 无线传感器在83%的测量中具有中等至良好可靠性。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: In pursuit to improve soft tissue balancing in total knee arthroplasties (TKAs), a wireless device was introduced to assess femorotibial pressures. The aim of this study was to evaluate the reliability of this device. METHODS: After 33 TKAs were balanced by conventional techniques, contact pressures were measured using a wireless sensor 3 times in a row; twice while the examiner was blinded for the result (n = 29); and once while the examiner was able to see the result as visual feedback (n = 32). Femorotibial pressures were measured in the medial and lateral compartments with the knee in 10°, 45°, and 90° of flexion (6 measurements per TKA). Furthermore, both the combined pressure and the difference in pressure between the compartments was calculated throughout the 3 positions (together another 6 measurements per TKA). RESULTS: The intraclass correlation coefficient between the blind measurements was poor in 2 of the 12 (17%), moderate in 4 of 12 (33%), and good in 6 of 12 (50%) measurements. The intraclass correlation coefficient between the blind and observing measurement was poor in 2 of the 12 (17%), moderate in 6 of 12 (50%), and good in 4 of 12 (33%) measurements. Especially measurements in 10° of flexion are associated with poorer reliability. CONCLUSION: The wireless sensor has a moderate to good reliability in 83% of the measurements.
引用
- van der Linde JA, Beath KJ, Leong AKL, et al. The Reliability of Sensor-Assisted Soft Tissue Measurements in Primary Total Knee Arthroplasty. J Arthroplasty. 2018. PMID: 29748068. doi:10.1016/j.arth.2018.03.067