单次使用患者特异性器械能否可靠用于初次全膝关节置换术?一项多中心对照研究
这项多中心对照研究评估了初次全膝关节置换术使用单次使用全一次性患者特异性截骨导板(SU)的临床与影像学结果,并与传统患者特异性截骨导板(PSI)和常规器械(CI)比较。70例连续患者使用SU,与140例历史患者(随机分配至PSI或CI)对比。结果显示三组在机械轴、临床评分、手术时间、输血需求和住院时长上无显著差异。
要点
- SU组机械轴平均179.8°(标准差3.1°),PSI组179.2°(标准差2.9°),CI组178.3°(标准差2.5°)。
- 三组机械轴离群值比例分别为CI组16/65(24.6%)、PSI组15/67(22.4%)、SU组14/70(20.0%),P=.81。
- 临床评分、手术时间、输血单位数和住院时长三组间无显著差异。
核心结论
- 在初次全膝关节置换术中使用单次使用患者特异性器械(SU)获得的临床和影像学结果与传统患者特异性截骨导板(PSI)和常规器械(CI)相似。
- 单次使用器械在初次全膝关节置换术中的潜在经济优势需进一步研究。
研究设计与样本
- 多中心对照研究:70例连续患者使用SU行初次全膝关节置换术,与140例历史患者(随机分配至PSI或CI)比较。
主要发现
- 机械轴(髋-膝-踝角)平均值:CI组179.8°(标准差3.1°),PSI组179.2°(标准差2.9°),SU组178.3°(标准差2.5°),P=.0082。
- 离群值:CI组16/65(24.6%),PSI组15/67(22.4%),SU组14/70(20.0%),P=.81。
- 临床结果无显著差异(P=.29和.19)。
- 手术时间、输血单位数和住院时长三组间无显著差异(P=.45、.31和.98)。
临床怎么用
- 原文未据此给出推荐。
局限性
- 单次使用器械在初次全膝关节置换术中的潜在经济优势需进一步研究。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The aim of this controlled multicenter study is to evaluate the clinical and radiologic outcomes of primary total knee arthroplasty (TKA) using single-use fully disposable and patient-specific cutting guides (SU) and compare the results to those obtained with traditional patient-specific cutting guides (PSI) vs conventional instrumentation (CI). METHODS: Seventy consecutive patients had their TKA performed using SU. They were compared to 140 historical patients requiring TKA that were randomized to have the procedure performed using PSI vs CI. The primary measure outcome was mechanical axis as measured on a standing long-leg radiograph using the hip-knee-ankle angle. Secondary outcome measures were Knee Society and Oxford knee scores, operative time, need for postoperative transfusion, and length of hospital stay. RESULTS: The mean hip-knee-ankle value was 179.8° (standard deviation [SD] 3.1°), 179.2° (SD 2.9°), and 178.3° (SD 2.5°) in the CI, PSI and SU groups, respectively (P = .0082). Outliers were identified in 16 of 65 (24.6%), 15 of 67 (22.4%), and 14 of 70 (20.0%) knees in the CI, PSI, and SU group, respectively (P = .81). There was no significant difference in the clinical results (P = .29 and .19, respectively). Operative time, number of unit transfusion, and length of hospital stay were not significantly different between the 3 groups (P = .45, .31, and 0.98, respectively). CONCLUSION: The use of an SU in TKA provided similar clinical and radiologic results to those obtained with traditional PSI and CI. The potential economic advantages of single-use instrumentation in primary TKA require further investigation.
引用
- Abane L, Zaoui A, Anract P, et al. Can a Single-Use and Patient-Specific Instrumentation Be Reliably Used in Primary Total Knee Arthroplasty? A Multicenter Controlled Study. J Arthroplasty. 2018. PMID: 29576488. doi:10.1016/j.arth.2018.02.038
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。