文献速递关节外科

金属对聚乙烯全髋关节置换术相比金属对金属全髋关节置换术和表面置换髋关节成形术具有更高比例的混合型或实性假肿瘤

Higher Prevalence of Mixed or Solid Pseudotumors in Metal-on-Polyethylene Total Hip Arthroplasty Compared With Metal-on-Metal Total Hip Arthroplasty and Resurfacing Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2018-02-12

该横断面研究纳入111例患者共148个髋关节(金属对金属全髋30个、金属对金属表面置换47个、金属对聚乙烯全髋71个),术后平均7.1年评估假肿瘤患病率。金属对聚乙烯全髋的混合型或实性假肿瘤比例显著高于金属对金属全髋和表面置换(P=0.01),且混合型或实性假肿瘤患者Harris髋关节评分和Oxford髋关节评分更差、钴离子水平更高。

要点

  • 金属对聚乙烯全髋假肿瘤患病率41%,与金属对金属全髋43%和表面置换28%相似(P=0.10)。
  • 金属对聚乙烯全髋混合型或实性假肿瘤10例,高于金属对金属全髋3例和表面置换0例(P=0.01)。
  • 混合型或实性假肿瘤患者Harris髋关节评分(P=0.01)和Oxford髋关节评分(P=0.002)更差,钴离子水平更高(P=0.0009)。

核心结论

  • 金属对聚乙烯全髋的假肿瘤患病率与金属对金属全髋和表面置换相似。
  • 混合型或实性假肿瘤在金属对聚乙烯全髋中比金属对金属髋关节更常见。
  • 混合型或实性假肿瘤患者临床评分更差且钴离子水平更高。

研究设计与样本

  • 横断面研究,纳入111例患者共148个髋关节,包括30个金属对金属全髋、47个金属对金属表面置换和71个金属对聚乙烯全髋。
  • 术后平均7.1年(范围0.2-21.5年)进行金属伪影减少序列磁共振成像、金属离子测量、Harris髋关节评分、Oxford髋关节评分、Copenhagen髋和腹股沟结局评分及常规X线检查。

主要发现

  • 假肿瘤患病率:金属对金属全髋13/30(43%)、金属对金属表面置换13/47(28%)、金属对聚乙烯全髋29/71(41%),组间相似(P=0.10)。
  • 混合型或实性假肿瘤患病率:金属对聚乙烯全髋10例,高于金属对金属全髋3例和金属对金属表面置换0例(P=0.01)。
  • 混合型或实性假肿瘤髋的Harris髋关节评分(P=0.01)和Oxford髋关节评分(P=0.002)显著更差,钴离子水平显著更高(P=0.0009), compared to 无假肿瘤或囊性假肿瘤髋。

关键数值与推荐

  • 假肿瘤患病率组间比较P=0.10。
  • 混合型或实性假肿瘤患病率组间比较P=0.01。
  • 混合型或实性假肿瘤与Harris髋关节评分P=0.01、Oxford髋关节评分P=0.002、钴离子水平P=0.0009相关。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Pseudotumors are a common finding in metal-on-metal (MoM) total hip arthroplasty (THA) and resurfacing hip arthroplasty (RHA). However, information on pseudotumors in metal-on-polyethylene (MoP) THA is limited. METHODS: One hundred eleven patients with 148 hip articulations-30 MoM THA, 47 MoM RHA, and 71 MoP THA-participated in a cross-sectional study at mean 7.1 (range: 0.2-21.5) years postoperatively. Patients were evaluated with metal artifact reducing sequence magnetic resonance imaging, measurements of metal ions, clinical scores of Harris Hip Score, Oxford Hip Score, the Copenhagen Hip and Groin Outcome Score, and conventional radiographs. RESULTS: Pseudotumors were present in 13 of 30 (43%) MoM THA, 13 of 47 (28%) MoM RHA, and 29 of 71 (41%) MoP THA patients, which was a similar prevalence (P = .10). The prevalence of mixed or solid pseudotumors was significantly higher in patients with MoP THA (n = 10) compared to MoM THA (n = 3) and MoM THA (n = 0), (P = .01). Hips with a mixed or solid pseudotumor had significantly poorer scores of Harris Hip Score (P = .01) and OHS (P = .002) and higher metal ion levels of cobalt (P = .0009) compared to hips without a pseudotumor or with a cystic pseudotumor. CONCLUSION: Pseudotumors have primarily been associated with MoM hip articulations, but we found a similar pseudotumor prevalence in MoP THA, which is the most common bearing worldwide. Mixed or solid pseudotumors were more often seen in MoP THA compared with MoM hip articulations, and patients with a mixed or solid pseudotumor had poorer clinical scores and higher metal ion levels than patients without a pseudotumor or with a cystic pseudotumor.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Hjorth MH, Mechlenburg I, Soballe K, et al. Higher Prevalence of Mixed or Solid Pseudotumors in Metal-on-Polyethylene Total Hip Arthroplasty Compared With Metal-on-Metal Total Hip Arthroplasty and Resurfacing Hip Arthroplasty. J Arthroplasty. 2018. PMID: 29526334. doi:10.1016/j.arth.2018.02.011

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。