第二代电子韧带平衡装置在全膝关节置换术中的应用:一项尸体研究
研究在12具尸体膝关节上比较第二代电子韧带平衡装置与2种机械平衡器械的准确性,并依次在TKA术前、安装试模股骨假体后、内侧松解后及切断后交叉韧带后测量软组织平衡。电子装置在测量分离间隙和分离力方面比机械器械更准确,内侧松解增加屈曲位内侧间隙及被动膝外翻松弛度,后交叉韧带切断增加屈伸位胫股间隙且外侧间隙增加更大。
要点
- 电子平衡装置在测量分离间隙和分离力方面比机械器械更准确。
- TKA术前平均屈曲间隙宽于伸直间隙,内侧间隙比外侧间隙更紧。
- 内侧松解增加屈曲位内侧间隙和被动膝外翻松弛度;后交叉韧带切断增加屈伸位胫股间隙,外侧间隙增加更大。
核心结论
- 第二代电子平衡装置显著比机械器械更准确,并能记录整个屈曲范围内的膝关节平衡。
- 更准确的软组织平衡可能改善TKA术后结局。
研究设计与样本
- 在12具尸体膝关节上实施TKA,比较第二代电子韧带平衡装置与2种机械平衡器械的准确性。
- 依次在TKA术前、安装试模股骨假体后、内侧松解后及切断后交叉韧带后测量软组织平衡。
- 在内侧松解前后测量10 Nm外翻力矩下的膝关节冠状面松弛度。
主要发现
- 电子平衡器械在测量分离间隙和分离力方面比机械器械更准确。
- 平均而言,TKA术前屈曲间隙宽于伸直间隙,内侧间隙比外侧间隙更紧。
- 内侧松解增加屈曲位内侧间隙并增加被动膝外翻松弛度。
- 后交叉韧带切断增加屈伸位胫股间隙,且外侧间隙增加更大。
关键数值与推荐
- 冠状面松弛度在10 Nm外翻力矩下测量。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Knee instability is emerging as a major complication after total knee arthroplasty (TKA), with ligament laxity and component alignment listed as important contributory factors. Knee balancing remains an art and is largely dependent on the surgeon's subjective "feel." The objectives were to measure the accuracy of an electronic balancing device to document the magnitude of correction in knee balance after soft-tissue releases and measure change in knee laxity after medial release. METHODS: The accuracy of a second-generation electronic ligament-balancing device was compared with that of 2 mechanical balancing instruments. TKA was performed in 12 cadaver knees. Soft-tissue balance was measured sequentially before TKA, after mounting a trial femoral component, after medial release, and after resecting the posterior cruciate ligament. Coronal laxity of the knee under a 10 Nm valgus moment was measured before and after medial release. RESULTS: The electronic balancing instrument was more accurate than mechanical instruments in measuring distracted gap and distraction force. On average, before TKA, the flexion gap was wider than the extension gap, and the medial gap was tighter than the lateral gap. Medial release increased the medial gap in flexion and increased passive knee valgus laxity. Posterior cruciate ligament release increased the tibiofemoral gap in both flexion and extension with a greater increase in the lateral gap. CONCLUSION: The second-generation electronic balancing device was significantly more accurate than mechanical instruments and could record knee balance over the entire range of flexion. More accurate soft-tissue balance may enhance outcomes after TKA.
引用
- Nielsen ES, Hsu A, Patil S, et al. Second-Generation Electronic Ligament Balancing for Knee Arthroplasty: A Cadaver Study. J Arthroplasty. 2018. PMID: 29555494. doi:10.1016/j.arth.2018.02.057
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。