单一设计大直径头模块化金属对金属全髋关节假体的回收分析
本研究对来自单一制造商的24例回收模块化金属对金属全髋植入物进行综合分析,评估轴承表面损伤、磨损及模块化锥度腐蚀,并与患者、手术和植入物特征关联。结果显示,金属对金属相关翻修患者的血金属离子水平显著更高,轴承磨损与血金属离子水平强相关,且在髋臼杯放置更垂直时显著增加。年轻患者体质量指数更高且髋臼杯放置更差。
要点
- 金属对金属相关翻修患者血金属离子水平显著更高。
- 轴承磨损与血金属离子水平强相关,髋臼杯更垂直时磨损显著增加。
- 年轻患者体质量指数更高且髋臼杯放置更差。
核心结论
- 髋臼杯倾斜角被认为是导致翻修手术的主要原因,通过增加金属对金属磨损、升高血金属离子水平及后续不良局部组织反应实现。
- 体质量较重的患者可能增加手术难度,本队列中显示与髋臼杯放置不佳相关。
研究设计与样本
- 对来自单一制造商的24例回收模块化金属对金属全髋植入物进行回收分析,比较轴承表面损伤、磨损和模块化锥度腐蚀与患者、手术及植入物特征。
- 收集临床、患者和手术数据包括年龄、体质量指数、血金属离子水平和髋臼杯倾斜度;采用视觉评估和表面轮廓术进行损伤评估;测量髋臼内衬和股骨头的体积磨损;测量股骨头锥度孔因腐蚀和微动导致的材料去除。
主要发现
- 金属对金属相关翻修原因的患者血金属离子水平显著更高。
- 轴承磨损与血金属离子水平强相关,并且在髋臼杯放置更垂直时显著增加。
- 年轻患者倾向于有更高的体质量指数以及更差的髋臼杯放置。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: There are limited publications examining modular metal-on-metal (MoM) total hip implants in which a comprehensive analysis of retrieved components is performed. This study examines 24 retrieved modular MoM implants from a single manufacturer and compares retrieval analytics; bearing surface damage, wear, and modular taper corrosion against patient, surgical and implant characteristics to elucidate significant associations. METHODS: Clinical, patient, and surgical data were collected including age, body mass index, blood metal ion levels, and cup inclination. Damage assessment was performed visually in addition to surface profilometry. Acetabular liners and femoral heads were measured for volumetric wear. Femoral head taper bores were similarly measured for material removal due to corrosion and fretting. RESULTS: Patients with MoM-related reasons for revision showed significantly higher levels of blood metal ion levels. Bearing wear was strongly associated with blood metal ion levels and was significantly increased in cups placed more vertically. Younger patients tended to have higher body mass indices as well as poorer cup placement. CONCLUSION: This work details a broad range of analyses on a series of modular MoM total hip implants from a single manufacturer of which there are few published studies. Acetabular cup inclination angle was deemed a primary cause of revision surgery through increased MoM wear, high metal ion levels in the blood, and subsequent adverse local tissue reactions. Heavy patients can increase the surgical difficulty which was shown to be related to poor cup placement in this cohort.
引用
- Gascoyne TC, Turgeon TR, Burnell CD, et al. Retrieval Analysis of Large-Head Modular Metal-on-Metal Hip Replacements of a Single Design. J Arthroplasty. 2018. PMID: 29402714. doi:10.1016/j.arth.2017.12.044
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。