全髋关节置换术中的数字放射摄影:技术与放射学结果
研究前瞻性纳入369例连续接受全髋关节置换术的患者,术中使用数字放射摄影评估髋臼杯位置,目标为外展30°至50°、前倾15°至35°。结果显示97.8%的髋臼杯置于外展30°至50°范围内(平均39.5°±4.6°),97.6%置于前倾15°至35°范围内(平均26.6°±4.7°),28%的髋臼杯根据术中测量重新定位。
要点
- 97.8%的髋臼杯外展角度在30°至50°目标范围内,平均39.5°±4.6°。
- 97.6%的髋臼杯前倾角度在15°至35°目标范围内,平均26.6°±4.7°。
- 28%的髋臼杯根据术中测量重新定位,3%的髋关节在活动度测试中出现半脱位。
核心结论
- 使用数字放射摄影可将97.8%的髋臼杯置于目标区域,2.2%有意置于目标区域外。
- 结合严格撞击测试,数字放射摄影可使全髋关节置换术中的髋臼杯放置具有可预测性。
研究设计与样本
- 前瞻性纳入369例连续接受全髋关节置换术的患者。
- 术前拍摄仰卧位骨盆正位X线片,术中在侧卧位下拍摄骨盆正位X线片,调整射线角度和手术床以近似术前片的旋转和倾斜。
- 髋臼杯位置目标为外展30°至50°、前倾15°至35°,术中测量计算并在严格撞击和活动度测试后确定最终位置,术后拍摄骨盆正位X线片,由两名独立观察者重新测量所有外展和前倾角度。
主要发现
- 97.8%的髋臼杯置于外展30°至50°范围内,平均角度为39.5°±4.6°。
- 2.2%的髋臼杯置于目标区域外,是基于术中活动度测试和患者因素有意为之。
- 97.6%的髋臼杯置于前倾15°至35°范围内,平均角度为26.6°±4.7°。
- 28%的髋臼杯根据术中测量重新定位。
- 3%的髋关节在活动度测试中出现半脱位,尽管测量结果可接受,仍需重新定位髋臼杯;另有1例早期前脱位。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Obtaining the ideal acetabular cup position in total hip arthroplasty remains a challenge. Advancements in digital radiography and image analysis software allow the assessment of the cup position during the surgical procedure. This study describes a validated technique for evaluating cup position during total hip arthroplasty using digital radiography. METHODS: Three hundred and sixty-nine consecutive patients undergoing total hip arthroplasty were prospectively enrolled. Preoperative supine anteroposterior pelvic radiographs were made. Intraoperative anteroposterior pelvic radiographs were made with the patient in the lateral decubitus position. Radiographic beam angle adjustments and operative table adjustments were made to approximate rotation and tilt of the preoperative radiograph. The target for cup position was 30° to 50° abduction and 15° to 35° anteversion. Intraoperative radiographic measurements were calculated and final cup position was determined after strict impingement and range-of-motion testing. Postoperative anteroposterior pelvic radiographs were made. Two independent observers remeasured all abduction and anteversion angles. RESULTS: Of the cups, 97.8% were placed within 30° to 50° of abduction, with a mean angle (and standard deviation) of 39.5° ± 4.6°. The 2.2% of cups placed outside the target zone were placed so purposefully on the basis of intraoperative range-of-motion testing and patient factors, and 97.6% of cups were placed between 15° and 35° of anteversion, with a mean angle of 26.6° ± 4.7°. Twenty-eight percent of cups were repositioned on the basis of intraoperative measurements. Subluxation during range-of-motion testing occurred in 3% of hips despite acceptable measurements, necessitating cup repositioning. There was 1 early anterior dislocation. CONCLUSIONS: Placing the acetabular component within a target range is a critical component to minimizing dislocation and polyethylene wear in total hip arthroplasty. Using digital radiography, we positioned the acetabular component in our desired target zone in 97.8% of cases and outside the target zone, purposefully, in 2.2% of cases. When used in conjunction with strict impingement testing, digital radiography allows for predictable cup placement in total hip arthroplasty.
引用
- Penenberg BL, Samagh SP, Rajaee SS, et al. Digital Radiography in Total Hip Arthroplasty: Technique and Radiographic Results. J Bone Joint Surg Am. 2018. PMID: 29406344. doi:10.2106/JBJS.16.01501
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。