回收的金属对聚乙烯模块化全髋关节置换系统中微动与腐蚀损伤:股骨头尺寸的重要性

Fretting and Corrosion Damage in Retrieved Metal-on-Polyethylene Modular Total Hip Arthroplasty Systems: What Is the Importance of Femoral Head Size?
PubMed 文献 · J Arthroplasty · 2017-10-16

研究评估了92例回收的初次金属对聚乙烯模块化全髋关节置换假体,对锥度界面(股骨柄锥和股骨头)的微动、腐蚀和损伤进行分级,并分析假体设计、影像学因素和患者因素与损伤的关联。结果显示较大的股骨头尺寸与较轻的股骨头腐蚀和微动相关,28毫米股骨头表现出更多中重度损伤。

要点

  • 男性患者股骨头腐蚀更严重(P=0.037)。
  • 股骨头尺寸与股骨头微动(ρ=-0.26,P=0.007)和腐蚀(ρ=-0.21,P=0.028)呈弱负相关。
  • 股骨柄内翻对齐与更严重的股骨头微动相关(P=0.038)。

核心结论

  • 较大的股骨头尺寸与较轻的股骨头腐蚀和股骨头微动相关,28毫米股骨头表现出更多中重度损伤。
  • 其他因素如股骨头-锥度接合和几何形状而非股骨头尺寸可能影响锥度界面的腐蚀和微动损伤率。

研究设计与样本

  • 研究评估了92例回收的初次金属对聚乙烯模块化全髋关节置换假体,对锥度界面(包括股骨柄锥和股骨头)的微动、腐蚀和损伤进行分级。
  • 术前X线片评估骨溶解和股骨柄对齐,并查阅病历获取人口统计学数据。

主要发现

  • 男性患者股骨头腐蚀更严重(P=0.037)。
  • 患者翻修时年龄与柄锥腐蚀呈弱负相关(ρ=-0.20,P=0.04)。
  • 体重指数与股骨头微动呈弱正相关(ρ=0.26,P=0.01),植入时间与股骨头微动呈弱正相关(ρ=0.33,P=0.001)。
  • 股骨头尺寸与股骨头微动呈弱负相关(ρ=-0.26,P=0.007),与股骨头腐蚀呈弱负相关(ρ=-0.21,P=0.028)。
  • 股骨头偏移量与柄锥微动呈弱正相关(ρ=0.23,P=0.030),股骨柄内翻对齐与更严重的股骨头微动相关(P=0.038)。

关键数值与推荐

  • 男性患者股骨头腐蚀更严重(P=0.037)。
  • 患者翻修时年龄与柄锥腐蚀呈弱负相关(ρ=-0.20,P=0.04)。
  • 体重指数与股骨头微动呈弱正相关(ρ=0.26,P=0.01),植入时间与股骨头微动呈弱正相关(ρ=0.33,P=0.001)。
  • 股骨头尺寸与股骨头微动呈弱负相关(ρ=-0.26,P=0.007),与股骨头腐蚀呈弱负相关(ρ=-0.21,P=0.028)。
  • 股骨头偏移量与柄锥微动呈弱正相关(ρ=0.23,P=0.030),股骨柄内翻对齐与更严重的股骨头微动相关(P=0.038)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Fretting and corrosion at the modular femoral head-femoral neck (taper) interface have been reported in retrieved total hip arthroplasty (THA) prostheses. This study investigated associations among implant design, radiographic factors, and patient factors with corrosion and fretting at the taper interface in retrieved metal-on-polyethylene modular THA prostheses. METHODS: Ninety-two retrieved primary metal-on-polyethylene THA implants were evaluated and graded for fretting, corrosion, and damage at the taper interface, including the femoral stem trunnion and femoral head. Preoperative radiographs were assessed for osteolysis and femoral stem alignment; and medical records were reviewed for demographic data. RESULTS: Male patients had greater head corrosion (P = .037), patient age at revision had a weak, negative correlation with trunnion corrosion (ρ = -0.20, P = .04), and both body mass index and duration of implantation had weak, positive correlations with head fretting (ρ = 0.26, P = .01 and ρ = 0.33, P = .001, respectively). A weak, negative correlation was found between femoral head size and both head fretting and head corrosion (ρ = -0.26, P = .007 and ρ = -0.21, P = .028, respectively), and a weak, positive correlation was found between head offset and trunnion fretting (ρ = 0.23, P = .030). Varus femoral stem alignment was associated with greater head fretting (P = .038). CONCLUSION: Larger femoral head sizes were correlated with less severe head corrosion and head fretting, with 28-mm heads exhibiting more moderate-to-severe damage. Other factors, such as head-taper engagement and geometry, rather than head size, may affect rates of corrosion and fretting damage at the taper interface.