文献速递关节外科

R3髋臼杯在常规摩擦界面中未显示高失败率:至少5年随访

R3 Cup Does Not Have a High Failure Rate in Conventional Bearings: A Minimum of 5-Year Follow-Up.
PubMed 文献 · J Arthroplasty · 2017-10-07

本研究回顾性分析本中心293例连续全髋关节置换(286例患者)中使用R3髋臼杯的病例,主要结局为翻修,次要结局为牛津髋关节评分(OHS)和影像学评估。平均年龄69.4岁,术前平均OHS 23,末次随访平均OHS 40,5年翻修风险为1.11%,所有患者5年时ARA评分优秀,末次随访无R3杯周围骨溶解。作者认为R3髋臼系统配合常规摩擦界面具有高生存率,与2015年英国骨科数据评估小组(ODEP)5A*评级一致。

要点

  • 293例连续全髋关节置换使用R3髋臼杯,5年翻修风险为1.11%,仅2例R3杯翻修。
  • 术前平均OHS 23,末次随访平均OHS 40,5年时所有患者ARA评分优秀。
  • 末次随访无R3杯周围骨溶解,作者认为生存率与ODEP 5A*评级一致。

核心结论

  • 作者认为,使用R3髋臼系统配合常规摩擦界面的经验显示高生存率,与2015年英国ODEP 5A*评级一致。

研究设计与样本

  • 回顾性分析本中心关节置换数据库中使用R3髋臼杯的293例连续全髋关节置换(286例患者),至少5年随访,主要结局为翻修。

主要发现

  • 患者平均年龄69.4岁,术前平均OHS为23(范围10-34),末次随访平均OHS为40(范围33-48)。
  • 5年时所有患者影像学评估ARA评分优秀,末次随访无R3杯周围骨溶解。
  • 本系列共3例翻修,其中2例R3杯翻修,5年翻修风险为1.11%。

关键数值与推荐

  • 5年翻修风险为1.11%,共3例翻修,其中2例为R3杯翻修。
  • 术前平均OHS 23(范围10-34) vs 末次随访平均OHS 40(范围33-48)。
  • 5年时所有患者ARA评分优秀,末次随访无R3杯周围骨溶解。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The R3 cementless acetabular system was first marketed in Australia and Europe in 2007. Previous papers have shown high failure rates of the R3 cup with up to 24% with metal-on-metal bearing. There are currently no medium term clinical results on this cup. The aim of the study is to review our results of the R3 acetabular cup with conventional bearings with a minimum of 5-year follow-up. METHODS: Patients who were implanted with the R3 acetabular cup were identified from our center's arthroplasty database. A total of 293 consecutive total hip arthroplasties were performed in 286 patients. The primary outcome was revision. The secondary outcomes were the Oxford Hip Scores (OHS) and radiographic evaluation. RESULTS: The mean age of the patients was 69.4 years. The mean preoperative OHS was 23 (range 10-34) and the mean OHS was 40 (range 33-48) at the final follow-up. Radiological evaluation showed an excellent ARA score in all patients at 5 years. None of the R3 cups showed osteolysis at the final follow-up. There were 3 revisions in our series, of which 2 R3 cups were revised. The risk of revision was 1.11% at 5 years. CONCLUSION: Our experience of using the R3 acetabular system with conventional bearings showed high survivorship and is consistent with the allocated Orthopaedic Data Evaluation Panel rating of 5A* as rated in 2015 in the United Kingdom.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Teoh KH, Whitham RDJ, Golding DM, et al. R3 Cup Does Not Have a High Failure Rate in Conventional Bearings: A Minimum of 5-Year Follow-Up. J Arthroplasty. 2017. PMID: 29107497. doi:10.1016/j.arth.2017.09.059

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。