活动平台单髁膝关节置换术后轴承活动度减小与功能结局不良相关
本研究从150例连续牛津内侧UKA病例中,将术后1年功能结局一般/差(OKS<34,n=15)的患者与功能结局良好/优(OKS≥34,n=15)的患者按年龄、性别、术前临床评分和随访期匹配,在透视下进行踏步和深屈膝运动的在体运动学评估。结果发现两组胫骨后倾、完成运动时间和髌腱角趋势相似,但OKS<34组轴承总活动度较OKS≥34组显著减小(35%),且轴承在胫骨上平均更靠后1.7 mm。结论认为异常膝关节运动学,特别是轴承活动和位置,与活动平台UKA术后功能结局较差相关。
要点
- OKS<34组轴承总活动度较OKS≥34组显著减小35%。
- OKS<34组轴承在胫骨上平均较OKS≥34组更靠后1.7 mm。
- 两组胫骨后倾、完成运动时间和髌腱角趋势相似。
核心结论
- 本研究提供证据表明异常膝关节运动学,特别是轴承活动和位置,与活动平台UKA术后功能结局较差相关。
研究设计与样本
- 从150例连续牛津内侧UKA病例中识别术后1年功能结局一般/差(OKS<34,n=15)的患者,并按年龄、性别、术前临床评分和随访期与功能结局良好/优(OKS≥34,n=15)的患者匹配。
- 在透视成像下进行踏步和深屈膝运动的在体运动学评估,使用MATLAB软件分析透视视频以测量完成运动所需时间的变化、髌腱角和轴承位置随膝关节屈曲角的变化。
主要发现
- 功能结局一般/差组平均OKS为29.9,功能结局良好/优组平均OKS为41.1。
- 两组胫骨后倾、完成运动所需时间和髌腱角随屈曲范围的变化趋势相似。
- OKS<34组轴承总活动度较OKS≥34组显著减小35%。
- OKS<34组轴承在胫骨上平均较OKS≥34组更靠后1.7 mm。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: A small proportion of patients with mobile unicompartmental knee arthroplasty (UKA) report poor functional outcomes in spite of optimal component alignment on postoperative radiographs. The purpose of this study is to assess whether there is a correlation between functional outcome and knee kinematics. METHODS: From a cohort of consecutive cases of 150 Oxford medial UKA, patients with fair/poor functional outcome at 1-year postsurgery (Oxford Knee Score [OKS] < 34, n = 15) were identified and matched for age, gender, preoperative clinical scores, and follow-up period with a cohort of patients with good/excellent outcome (OKS ≥ 34, n = 15). In vivo kinematic assessment was performed using step-up and deep knee bend exercises under fluoroscopic imaging. The fluoroscopic videos were analyzed using MATLAB software to measure the variation in time taken to complete the exercises, patellar tendon angle, and bearing position with knee flexion angle. RESULTS: Mean OKS in the fair/poor group was 29.9 and the mean OKS in the good/excellent group was 41.1. The tibial slope, time taken to complete the exercises, and patellar tendon angle trend over the flexion range were similar in both the groups; however, bearing position and the extent of bearing excursion differed significantly. The total bearing excursion in the OKS < 34 group was significantly smaller than the OKS ≥ 34 group (35%). Furthermore, on average, the bearing was positioned 1.7 mm more posterior on the tibia in the OKS < 34 group. CONCLUSION: This study provides evidence that abnormal knee kinematics, in particular bearing excursion and positioning, are associated with worse functional outcomes after mobile UKA.
引用
- Wahal N, Gaba S, Malhotra R, et al. Reduced Bearing Excursion After Mobile-Bearing Unicompartmental Knee Arthroplasty is Associated With Poor Functional Outcomes. J Arthroplasty. 2017. PMID: 29103778. doi:10.1016/j.arth.2017.09.057
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。