全髋关节置换术中头-柄组件的混用
研究通过扭矩脱离实验测试6家厂商的髋臼柄与金属股骨头,先按原配组合测试,再将柄与其他厂商的头组合。Link、Smith & Nephew和Zimmer的柄与其他厂商头组合时锥度连接强度无显著差异;Biomet柄与Aesculap、DePuy或Smith & Nephew的头组合时强度显著降低。作者认为混用可能影响锥度连接强度并降低稳定性,不推荐混用不同厂商的头和股骨柄。
要点
- Link、Smith & Nephew和Zimmer柄与其他厂商头组合时锥度连接强度无显著差异。
- Biomet柄与Aesculap、DePuy或Smith & Nephew头组合时锥度连接强度显著降低。
- 作者认为混用不同厂商的头和股骨柄可能影响锥度连接强度并降低稳定性,不推荐混用。
核心结论
- 作者认为混用不同厂商的组件可能影响锥度连接强度并降低稳定性,不推荐混用不同厂商的头和股骨柄。
研究设计与样本
- 采用扭矩脱离实验,使用6家不同植入物厂商的髋臼柄和金属股骨头,先按原配组合测试,再将柄与其他厂商的金属头组合测试。
主要发现
- Link、Smith & Nephew和Zimmer的柄与其他厂商的头组合时,锥度连接强度无显著差异。
- Biomet柄与Aesculap、DePuy或Smith & Nephew的股骨头组合时,锥度连接强度显著降低。
- DePuy柄与原配股骨头组合时,锥度连接强度显著低于与Link、Biomet和Zimmer的头组合。
- Aesculap柄与Zimmer头组合时也观察到锥度连接强度显著降低。
临床怎么用
- 作者认为混用不同厂商的头和股骨柄可能影响锥度连接强度并降低稳定性,不推荐混用。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Implant manufacturers proclaim that the tapers of modular total hip arthroplasty are not standardized and can vary from manufacturer to manufacturer. That is why the combination of various components from different manufacturers ("Mix and Match") is not permitted. In this study, different taper combinations were investigated experimentally to assess the effect of "Mix and Match" regarding the taper connection strength. METHODS: Torque-off tests using hip stems and metal femoral heads from 6 different implant manufacturers were performed. First the components were tested as intended and afterwards the stems were combined with metal heads from other manufacturers. RESULTS: There was no significant difference in taper connection strength when stems from the manufacturers Link, Smith & Nephew, and Zimmer were combined with heads from other manufacturers. The Biomet stems showed a significantly reduced taper connection strength if femoral heads of Aesculap, DePuy, or Smith & Nephew were used. On the contrary, the DePuy stems in combination with the originally intended femoral heads showed a significantly lower taper connection strength compared to the use of heads from Link, Biomet, and Zimmer. The same was observed for the Aesculap stems in combination with Zimmer heads. CONCLUSION: The results of this study suggest that mixing components from different manufacturers may affect the taper connection strength and could reduce the stability. As safety should be a high priority in patient treatment, any potential risks should be avoided. Therefore, mixing and matching of heads and femoral stems from different manufacturers cannot be recommended.
引用
- Mueller U, Panzram B, Braun S, et al. Mixing of Head-Stem Components in Total Hip Arthroplasty. J Arthroplasty. 2017. PMID: 29174405. doi:10.1016/j.arth.2017.10.047
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。