相似与混合金属头颈锥形连接处柄锥材料损失的比较研究

A comparison study of stem taper material loss at similar and mixed metal head-neck taper junctions.
PubMed 文献 · Bone Joint J · 2017-01-01

该研究通过取出的较大直径金属对金属全髋关节假体,比较钴铬合金与钛合金股骨柄锥形连接处的体积材料损失,发现调整其他因素后钴铬合金柄锥的材料损失显著大于钛合金柄锥。

要点

  • 取出的较大直径金属对金属髋关节中,钴铬合金柄锥体积材料损失显著大于钛合金柄锥。
  • 仅纳入粗糙男性锥面且标称夹角接近5.666°的12/14锥形连接结构。
  • 翻修术中若发现锥形连接处损伤,应怀疑标准钴铬合金男性锥体几何形状发生显著改变。

核心结论

  • 当较大直径金属对金属髋关节翻修术中发现锥形连接处损伤时,应怀疑由标准钴铬合金制成的男性锥体已发生显著的锥形几何形状改变,且可能比钛合金柄锥更广泛。

研究设计与样本

  • 对取出的较大直径金属对金属全髋关节假体进行取回物分析,使用钛合金股骨柄(n=28)或钴铬合金股骨柄(n=21)。
  • 仅纳入粗糙男性锥面且标称夹角接近5.666°的12/14锥形连接结构。
  • 使用坐标测量机、轮廓仪和扫描电子显微镜绘制材料损失,并以锥度角、锥面粗糙度、轴承直径(水平杠杆臂)为自变量进行多元回归建模。

主要发现

  • 调整其他因素后,钴铬合金柄锥的体积材料损失显著大于等效的钛合金柄锥。

临床怎么用

  • 较大直径金属对金属髋关节翻修术中若发现锥形连接处损伤,应怀疑标准钴铬合金男性锥体已发生显著的锥形几何形状改变,且可能比钛合金柄锥更广泛。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: We sought to determine whether cobalt-chromium alloy (CoCr) femoral stem tapers (trunnions) wear more than titanium (Ti) alloy stem tapers (trunnions) when used in a large diameter (LD) metal-on-metal (MoM) hip arthroplasty system. PATIENTS AND METHODS: We performed explant analysis using validated methodology to determine the volumetric material loss at the taper surfaces of explanted LD CoCr MoM hip arthroplasties used with either a Ti alloy (n = 28) or CoCr femoral stem (n = 21). Only 12/14 taper constructs with a rough male taper surface and a nominal included angle close to 5.666° were included. Multiple regression modelling was undertaken using taper angle, taper roughness, bearing diameter (horizontal lever arm) as independent variables. Material loss was mapped using a coordinate measuring machine, profilometry and scanning electron microscopy. RESULTS: After adjustment for other factors, CoCr stem tapers were found to have significantly greater volumetric material loss than the equivalent Ti stem tapers. CONCLUSION: When taper junction damage is identified during revision of a LD MoM hip, it should be suspected that a male taper composed of a standard CoCr alloy has sustained significant changes to the taper cone geometry which are likely to be more extensive than those affecting a Ti alloy stem. Cite this article: Bone Joint J 2017;99-B:1304-12.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Langton DJ, Sidaginamale RP, Joyce TJ, et al. A comparison study of stem taper material loss at similar and mixed metal head-neck taper junctions. Bone Joint J. 2017. PMID: 28963151. doi:10.1302/0301-620X.99B10.BJJ-2016-1005.R1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。