本研究对46例(57髋)平均年龄53岁、平均UCLA活动评分8分的年轻活跃患者行非骨水泥固定初次全髋关节置换,均使用28mm股骨头对退火高交联聚乙烯内衬。平均随访14年(范围10-17年),线性磨损率为0.03 mm/年(SD±0.02),无磨损相关并发症翻修,无骨溶解,假体生存率100%。

要点

  • 平均14年随访,线性磨损率0.03 mm/年(SD±0.02)。
  • 无磨损相关并发症翻修,无骨溶解,生存率100%。
  • WOMAC评分91/100,患者问卷评分5/100,HSS髋评分37/40。

核心结论

  • 第一代退火高交联聚乙烯在年轻活跃患者10-17年随访中显示优异的磨损和临床结果。
  • 退火高交联聚乙烯的氧化降解未对磨损产生不良影响,且无其他磨损相关并发症。

研究设计与样本

  • 对46例(57髋)平均年龄53岁、平均UCLA活动评分8分的患者行非骨水泥固定初次全髋关节置换,均使用28mm股骨头对高交联聚乙烯内衬。
  • 线性磨损率由2名独立观察者测量,并记录影像学和临床评估。

主要发现

  • 平均随访14年(范围10-17年),磨损率为0.03 mm/年(SD±0.02)。
  • 无磨损相关并发症翻修,以磨损相关并发症翻修为终点,生存率为100%,未观察到骨溶解。
  • WOMAC评分91/100,患者问卷评分5/100,HSS髋评分37/40。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Concerns persist regarding the oxidative stability of annealed highly cross-linked polyethylene (HXLPE) and hence its wear performance especially long term in young and active patients. The purpose of this study is to assess the wear rates and clinical outcomes of a first-generation annealed HXLPE. METHODS: Forty-six patients (57 hips) with noncemented fixation, a mean age of 53 years, and a mean University of California Los Angeles activity score of 8 underwent primary total hip arthroplasty. A 28-mm femoral head against HXLPE liner was used in all cases. Linear wear rates were measured by 2 independent observers. Radiological and clinical assessments were recorded. RESULTS: At a mean of 14 years (range 10-17), the wear rate was 0.03 mm/y (SD ±0.02). There were no revisions for wear-related complications. Using revision for wear-related complications as an endpoint, the survivorship was 100% and we observed no osteolysis. Western Ontario and McMaster Universities Osteoarthritis Index, patient administered questionnaire, and Hospital for Special Surgery hip scores were 91/100, 5/100, and 37/40 respectively. CONCLUSION: This study confirms that a first-generation annealed HXLPE shows excellent wear and clinical results at 10-17 years in young and active patients. Oxidative degradation of an annealed HXLPE does not adversely affect wear and shows no other wear-related complications.