全髋关节置换术后假体迁移与聚乙烯磨损的测量:无需标记珠和特殊X线片
本文描述了一种新的计算机辅助方法,利用患者骨骼与假体模型,从全髋关节置换术后X线片测量假体迁移和聚乙烯磨损,并进行验证与可靠性评估。验证采用模拟已知迁移和穿透的体模X线片,临床可行性在两名患者中评估。髋臼和股骨假体迁移的测量一致性界限分别为0.37 mm和0.33 mm,股骨头穿透的一致性界限为0.161 mm。
要点
- 髋臼假体迁移测量的一致性界限为0.37 mm,股骨假体为0.33 mm。
- 股骨头穿透测量的一致性界限为0.161 mm。
- 该方法无需特殊X线片即可测量假体迁移和聚乙烯磨损。
核心结论
- 假体迁移和聚乙烯磨损可以在无需特殊X线片的情况下测量,准确测量可能有助于更早预测全髋关节置换术后失败。
研究设计与样本
- 方法学验证研究:利用厂家提供的髋臼和股骨假体模型及基于单次CT扫描构建的骨盆和股骨模型,通过计算机程序从随访的正位和侧位X线片测量假体迁移和股骨头穿透。
- 验证使用模拟已知迁移和穿透的体模X线片,临床可行性在两名患者中评估。
主要发现
- 髋臼假体迁移测量的一致性界限为0.37 mm,股骨假体迁移测量的一致性界限为0.33 mm。
- 股骨头穿透测量的一致性界限为0.161 mm。
关键数值与推荐
- 髋臼假体迁移测量一致性界限0.37 mm,股骨假体迁移测量一致性界限0.33 mm,股骨头穿透测量一致性界限0.161 mm。
临床怎么用
- 原文未据此给出推荐。
局限性
- 临床可行性仅在两名患者中评估。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: This paper describes the methodology, validation and reliability of a new computer-assisted method which uses models of the patient's bones and the components to measure their migration and polyethylene wear from radiographs after total hip arthroplasty (THA). MATERIALS AND METHODS: Models of the patient's acetabular and femoral component obtained from the manufacturer and models of the patient's pelvis and femur built from a single computed tomography (CT) scan, are used by a computer program to measure the migration of the components and the penetration of the femoral head from anteroposterior and lateral radiographs taken at follow-up visits. The program simulates the radiographic setup and matches the position and orientation of the models to outlines of the pelvis, the acetabular and femoral component, and femur on radiographs. Changes in position and orientation reflect the migration of the components and the penetration of the femoral head. Validation was performed using radiographs of phantoms simulating known migration and penetration, and the clinical feasibility of measuring migration was assessed in two patients. RESULTS: Migration of the acetabular and femoral components can be measured with limits of agreement (LOA) of 0.37 mm and 0.33 mm, respectively. Penetration of the femoral head can be measured with LOA of 0.161 mm. CONCLUSION: The migration of components and polyethylene wear can be measured without needing specialised radiographs. Accurate measurement may allow earlier prediction of failure after THA. Cite this article: Bone Joint J 2017;99-B:1290-7.
引用
- Devane PA, Horne JG, Foley G, et al. Measuring the migration of the components and polyethylene wear after total hip arthroplasty: beads and specialised radiographs are not necessary. Bone Joint J. 2017. PMID: 28963149. doi:10.1302/0301-620X.99B10.BJJ-2016-1169.R1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。