文献速递关节外科

回收的双极半髋假体中的额外关节活动是否比全髋关节置换术减少锥部磨损

Does the Additional Articulation in Retrieved Bipolar Hemiarthroplasty Implants Decrease Trunnionosis Compared to Total Hip Arthroplasty?
PubMed 文献 · J Arthroplasty · 2017-08-31

本研究对同一制造商单一锥度设计的双极半髋(BH)与金属对聚乙烯全髋关节置换(MoP THA)回收假体进行磨损与腐蚀评分分析,比较28 mm股骨头颈锥部的磨损腐蚀情况。结果显示BH组股骨头锥部总体微动磨损低于MoP THA组(P=0.02),所有假体总腐蚀评分与植入时间正相关(ρ=0.54,P<0.02)。

要点

  • BH组股骨头锥部总体微动磨损评分低于MoP THA组(P=0.02)。
  • 所有回收假体总腐蚀评分与植入时间正相关(ρ=0.54,P<0.02)。
  • 28 mm头该锥度设计中,MoP THA微动磨损增加未伴随腐蚀增加。

核心结论

  • 与MoP THA相比,BH回收股骨头微动磨损损伤减轻。
  • BH假体的额外关节活动可能减少头颈锥部连接处产生的扭矩,从而减少微动。
  • 在本锥度设计的28 mm头中,MoP THA假体微动磨损增加与腐蚀增加无关。
  • BH或MoP THA假体植入时间越长,腐蚀损伤风险越大。

研究设计与样本

  • 对同一制造商单一锥度设计、体内植入至少2年的BH与MoP THA假体进行回收分析。
  • 比较15个28 mm直径BH股骨头及对应股骨柄与按体内时间和头长(28 mm,-3 mm至28 mm,+8 mm)匹配的MoP THA假体。
  • 在立体显微镜下完成微动磨损和腐蚀损伤评分。

主要发现

  • BH回收股骨头锥部总体微动磨损低于MoP THA回收假体(P=0.02)。
  • 所有回收假体总腐蚀评分与植入时间正相关(ρ=0.54,P<0.02)。

关键数值与推荐

  • BH组股骨头锥部总体微动磨损评分低于MoP THA组(P=0.02)。
  • 所有回收假体总腐蚀评分与植入时间正相关(ρ=0.54,P<0.02)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Trunnionosis at the modular head-neck taper interface in metal-on-polyethylene total hip arthroplasty (MoP THA) has been shown to occur, and represents a potential mode of MoP THA failure. The purpose of the present investigation is to elucidate differences in fretting and corrosion at the head-neck taper interface of prostheses retrieved from bipolar hemiarthroplasty (BH) and MoP THA. METHODS: A retrieval analysis of BH and MoP THA prostheses featuring a single taper design from a single manufacturer and in vivo for a minimum 2 years was performed. Fifteen femoral heads of 28-mm diameter and corresponding femoral stems retrieved from BH were compared with MoP THA implants matched based on time in vivo and head length (28 mm, -3 mm to 28 mm, +8 mm). Fretting and corrosion damage scoring was completed under stereomicroscopic visualization. RESULTS: Femoral head bore tapers retrieved from BH exhibited decreased overall fretting (P = .02), when compared to those retrieved from MoP THA. Total corrosion scores for all retrieved implants were positively correlated with implantation time (ρ = 0.54, P < .02). CONCLUSION: Femoral heads retrieved from BH exhibit decreased fretting damage compared to those retrieved from MoP THA. The added articulation in BH implants may decrease torque produced at the head-neck taper junction, thereby decreasing fretting. Increased fretting damage in implants from MoP THA is not associated with increased corrosion in 28-mm heads of this taper design. The longer a BH or MoP THA prosthesis is implanted, the greater the risk of damage due to corrosion.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Del Balso C, Teeter MG, Tan SC, et al. Does the Additional Articulation in Retrieved Bipolar Hemiarthroplasty Implants Decrease Trunnionosis Compared to Total Hip Arthroplasty? J Arthroplasty. 2017. PMID: 29033155. doi:10.1016/j.arth.2017.08.027

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。