顶椎区植入物密度对胸段青少年特发性脊柱侧凸三维矫形比整体植入物密度更重要:基于棒去旋转与整体椎体去旋转技术
该研究通过10例Lenke 1型青少年特发性脊柱侧凸患者特异性生物力学模型,计算机模拟140次后路内固定,比较参考螺钉模式(每融合节段两枚螺钉)与六种减少螺钉的模式。结果显示替代模式在冠状面和矢状面矫形与参考模式相差在4°以内,顶椎区螺钉密度从每节段一枚增至两枚改善顶椎旋转矫正百分比(r=0.887,P<0.05),参考模式骨-螺钉平均力为243N±54N,替代模式低11%至48%。
要点
- 替代螺钉模式在冠状面和矢状面的矫形与参考最大螺钉密度模式相差在4°以内。
- 顶椎区螺钉密度从每节段一枚增至两枚与顶椎旋转矫正百分比改善相关(r=0.887,P<0.05)。
- 参考模式骨-螺钉平均力为243N±54N,替代模式低11%至48%。
核心结论
- 与参考最大螺钉密度模式相比,替代螺钉模式在冠状面和矢状面允许相似的矫形。
- 顶椎旋转矫正与顶椎区螺钉密度强相关;当模拟整体椎体去旋转技术时,相同整体螺钉密度的不同螺钉模式间顶椎旋转矫正差异显著。
- 高螺钉密度倾向于过度约束固定节段并导致骨-螺钉界面力更高。
研究设计与样本
- 通过不同椎弓根螺钉模式的脊柱侧凸内固定数值模拟进行三维矫形与骨-螺钉力的生物力学分析。
- 为10例青少年特发性脊柱侧凸(Lenke 1型)病例创建患者特异性脊柱生物力学模型。
- 每个病例分别模拟参考螺钉模式(每融合节段两枚螺钉)和六种螺钉更少的替代模式,共计算模拟140次后路内固定。
- 模拟手术操作和模型定义在模拟间保持不变,仅螺钉数量和分布不同。
主要发现
- 替代螺钉模式在冠状面和矢状面的平均矫形与参考模式相差在4°以内。
- 顶椎区螺钉密度从每节段一枚增至两枚改善顶椎旋转矫正百分比(r=0.887,P<0.05)。
- 参考螺钉模式骨-螺钉平均力为243N±54N,替代螺钉模式低11%至48%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
STUDY DESIGN: Biomechanical analysis of 3D correction and bone-screw forces through numerical simulations of scoliosis instrumentation with different pedicle screw patterns. OBJECTIVE: To analyze the effect of different screw densities and distributions on 3D correction and bone-screw forces in adolescent idiopathic scoliosis (AIS) instrumentation. SUMMARY OF BACKGROUND DATA: Instrumentation constructs with various numbers of pedicle screws and patterns have been proposed for thoracic AIS instrumentation. However, systematic biomechanical studies have not yet been completed on the appropriate screw patterns for optimal 3D correction. METHODS: Patient-specific biomechanical models of the spine were created for 10 AIS cases (Lenke 1). For each case, surgical instrumentation patterns were computationally simulated using respectively a reference screw pattern (two screws per level fused) and six alternative screw patterns with fewer screws. Simulated surgical maneuvers and model definition were unchanged between simulations except the number and distribution of screws. 3D correction and bone-screw forces were compared. RESULTS: A total of 140 posterior instrumentations were computationally simulated. Mean corrections in the coronal and sagittal planes with alternative screw patterns were within 4° to the reference pattern. Increasing screw density in the apical region from one to two screws per level improved percent apical vertebral rotation (AVR) correction (r = 0.887, P < 0.05). Average bone-screw force associated with the reference screw pattern was 243N ± 54N and those with the alternative screw patterns were 11% to 48% lower. CONCLUSION: Compared with the reference maximal screw density pattern, alternative screw patterns allowed similar corrections in the coronal and sagittal planes. AVR correction was strongly correlated with screw density in the apical region; AVR correction varied significantly with screw patterns of the same overall screw density when an en bloc vertebral derotation technique was simulated. High screw density tended to overconstrain the instrumented spine and resulted in higher forces at the bone-screw interface. LEVEL OF EVIDENCE: N/A.
引用
- Delikaris A, Wang X, Boyer L, et al. Implant Density at the Apex Is More Important Than Overall Implant Density for 3D Correction in Thoracic Adolescent Idiopathic Scoliosis Using Rod Derotation and En Bloc Vertebral Derotation Technique. Spine (Phila Pa 1976). 2018. PMID: 29059123. doi:10.1097/BRS.0000000000002465