从仰卧位到站立位髋臼杯方位变化及其对高交联聚乙烯磨损的影响

Change in Acetabular Cup Orientation From Supine to Standing Position and Its Effect on Wear of Highly Crosslinked Polyethylene.
PubMed 文献 · J Arthroplasty · 2017-08-31

研究纳入38例全髋置换术后平均12.5年患者,采用放射立体测量法比较仰卧位与站立位髋臼杯方位及高交联聚乙烯三维磨损率。站立位前倾角较仰卧位平均增加12°(P<.0001),外展角平均增加2°(P=.001);两体位磨损率无差异(P=.093),且杯方位与磨损率无相关性。

要点

  • 站立位较仰卧位髋臼杯前倾角平均增加12°(仰卧15.1°±10.4° vs 站立27.2°±10.5°,P<.0001)。
  • 站立位较仰卧位髋臼杯外展角平均增加2°(仰卧44.4°±6.8° vs 站立46.3°±7.7°,P=.001)。
  • 仰卧位与站立位磨损率无差异(0.067±0.070 mm/y vs 0.073±0.074 mm/y,P=.093),两体位杯方位与磨损率均无相关性。

核心结论

  • 高交联聚乙烯是一种宽容的承重材料,尽管推荐遵循传统髋臼位置目标区,确保髋关节稳定性和考虑患者功能位也是髋臼位置需考虑的重要目标。

研究设计与样本

  • 研究纳入38例患者,在术后平均12.5年时招募,所有患者接受同一设计髋臼杯、高交联内衬和28 mm钴铬股骨头。
  • 患者接受仰卧位和站立位放射立体测量检查,X射线源和探测器定位以获取前后位和交叉侧位X线片,从X线片测量髋臼杯位置和三维磨损率,并记录每例患者的结果评分。

主要发现

  • 前倾角从仰卧位(15.1°±10.4°)到站立位(27.2°±10.5°)平均增加12°,差异有统计学意义(P<.0001)。
  • 外展角从仰卧位(44.4°±6.8°)到站立位(46.3°±7.7°)平均增加2°,差异有统计学意义(P=.001)。
  • 仰卧位磨损率(0.067±0.070 mm/y)与站立位磨损率(0.073±0.074 mm/y)之间无差异(P=.093)。
  • 两种体位下髋臼杯方位与磨损率之间均无相关性。

临床怎么用

  • 作者认为,尽管推荐遵循传统髋臼位置目标区,确保髋关节稳定性和考虑患者功能位也是髋臼位置需考虑的重要目标。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The purpose of this study is to measure acetabular cup position and wear of the highly crosslinked polyethylene liner in the supine and standing position for patients at a minimum of 10 years after the operation. METHODS: A total of 38 patients were recruited at a mean of 12.5 years after the operation. All patients received a single acetabular cup design with a highly crosslinked liner and a 28-mm cobalt-chromium femoral head. Patients underwent supine and standing radiostereometric examinations in which the X-ray sources and detectors were positioned to obtain an anterior-posterior and cross-table lateral radiograph. Acetabular cup position and the three-dimensional wear rate were measured from the radiographs, and outcome scores were recorded for each patient. RESULTS: Anteversion significantly increased (P < .0001) a mean of 12° from supine (15.1° ± 10.4°) to standing (27.2° ± 10.5°) position. Inclination also significantly increased (P = .001) a mean of 2° from supine (44.4° ± 6.8°) to standing (46.3° ± 7.7°) position. There was no difference (P = .093) in wear rate between supine (0.067 ± 0.070 mm/y) and standing (0.073 ± 0.074 mm/y) positions. There were no correlations between cup orientation and wear rate in either position. CONCLUSION: Highly crosslinked polyethylene is a forgiving bearing material. Although adherence to the traditional acetabular position target zone is recommended, ensuring hip stability and consideration of the patient's functional position are also important objectives to consider for the acetabular position.