维生素E浸润高交联聚乙烯五年磨损研究

Wear of Vitamin E-Infused Highly Cross-Linked Polyethylene at Five Years.
PubMed 文献 · J Bone Joint Surg Am · 2017-01-01

这项随机对照研究对63髋进行放射立体分析,比较维生素E浸润高交联聚乙烯(E1)与无维生素E高交联聚乙烯(ArComXL)全髋关节置换术后5年内股骨头穿透率。结果显示E1组近端穿透中位数0.13 mm低于ArComXL组的0.20 mm;2至5年间E1近端穿透率0.02 mm/年也低于ArComXL的0.04 mm/年,但两组均观察到穿透增加。

要点

  • E1组5年近端穿透中位数0.13 mm,低于ArComXL组的0.20 mm。
  • 2至5年E1近端穿透率中位数0.02 mm/年,低于ArComXL的0.04 mm/年。
  • 2至5年两组均观察到穿透增加,ArComXL穿透率更高。

核心结论

  • 2至5年间两组均观察到穿透增加,ArComXL组穿透率更高,导致5年时近端和总穿透均大于E1组。

研究设计与样本

  • 随机对照研究,对63髋进行放射立体分析,随访至5年,比较E1与ArComXL的穿透率。

主要发现

  • 整个观察期E1近端穿透中位数0.13 mm(均值0.11 mm,95%CI 0.08至0.14 mm),低于ArComXL的0.20 mm(均值0.22 mm,95%CI 0.17至0.26 mm)。
  • 2至5年E1近端穿透率中位数0.02 mm/年(均值0.01 mm/年,95%CI 0.01至0.02 mm/年),ArComXL为0.04 mm/年(均值0.04 mm/年,95%CI 0.03至0.05 mm/年)。
  • 2至5年E1总穿透率中位数0.04 mm/年(均值0.04 mm/年,95%CI 0.03至0.05 mm/年),ArComXL为0.07 mm/年(均值0.08 mm/年,95%CI 0.06至0.10 mm/年)。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: In an earlier study with a 2-year follow-up of uncemented cups, we had reported low femoral-head penetration of vitamin E-infused highly cross-linked polyethylene liners (E1) compared with highly cross-linked liners without vitamin E (ArComXL). We studied the penetration rate of E1 compared with that of ArComXL, with a focus on changes occurring between 2 and 5 years after total hip arthroplasty. METHODS: In this randomized controlled study, we performed radiostereometric analysis of the penetration rate up to 5 years in 63 hips. RESULTS: During the total period of observation, the median proximal penetration for E1, 0.13 mm (mean, 0.11 mm [95% confidence interval (CI), 0.08 to 0.14 mm]), was lower than that for ArComXL, 0.20 mm (mean, 0.22 mm [95% CI, 0.17 to 0.26 mm]). The median proximal penetration rate between 2 and 5 years was 0.02 mm/yr (mean, 0.01 mm/yr [95% CI, 0.01 to 0.02 mm/yr]) for E1 and 0.04 mm/yr (mean, 0.04 mm/yr [95% CI, 0.03 to 0.05 mm/yr) for ArComXL. The corresponding median total (i.e., 3-dimensional resultant) penetration rates were 0.04 mm/yr (mean, 0.04 mm/yr [95% CI, 0.03 to 0.05 mm/yr]) for E1 and 0.07 mm/yr (mean, 0.08 mm/yr [95% CI, 0.06 to 0.10 mm/yr]) for ArComXL. CONCLUSIONS: From years 2 to 5, we observed increased penetration in both groups. The penetration rate was higher for ArComXL, resulting in more proximal and total penetration at 5 years than for E1. LEVEL OF EVIDENCE: Therapeutic Level I. See Instructions for Authors for a complete description of levels of evidence.

来源:J Bone Joint Surg Am,查看原文。版权归原作者所有。

引用

  1. Shareghi B, Johanson PE, Kärrholm J, et al. Wear of Vitamin E-Infused Highly Cross-Linked Polyethylene at Five Years. J Bone Joint Surg Am. 2017. PMID: 28872526. doi:10.2106/JBJS.16.00691

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。