全髋关节置换术三维术前规划软件的观察者间与观察者内可靠性

Interobserver and Intraobserver Reliability of Three-Dimensional Preoperative Planning Software in Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2017-09-01

研究纳入46例(60髋)初次全髋关节置换术患者的术前CT,由6名骨科医生间隔4周进行两次三维模板规划,评估组件尺寸与对线的观察者间和观察者内可靠性。结果显示杯和柄尺寸的观察者间ICC分别为0.907和0.944,观察者内平均ICC分别为0.965和0.972;骨坏死组可靠性显著高于骨关节炎组,严重半脱位和股骨髓腔形态影响可靠性。

要点

  • 三维模板规划对全髋关节置换术组件尺寸与对线显示出极好的可靠性。
  • 骨坏死组的观察者间和观察者内可靠性显著高于骨关节炎组。
  • 骨关节炎组中严重半脱位髋的杯尺寸与对线可靠性下降。

核心结论

  • 基于CT的三维模板规划在全髋关节置换术组件尺寸和对线方面显示出极好的可靠性,但受累关节的畸形会影响术前规划的可靠性。

研究设计与样本

  • 选取46例(14男、32女)初次全髋关节置换术患者的60髋术前CT,由6名骨科医生间隔4周进行两次三维模板规划,评估观察者间和观察者内可靠性。

主要发现

  • 杯尺寸和柄尺寸的观察者间可靠性均为极好,ICC分别为0.907和0.944。
  • 观察者间可靠性在骨坏死组显著高于骨关节炎组。
  • 骨关节炎组中,严重半脱位髋的杯尺寸和对线可靠性下降。
  • 柄尺寸的百分比一致性在不同股骨髓腔形态间存在显著差异。
  • 杯尺寸的观察者内平均ICC总体为0.965,骨坏死组显著高于骨关节炎组;柄尺寸的观察者内平均ICC总体为0.972。

关键数值与推荐

  • 观察者间杯尺寸ICC=0.907,柄尺寸ICC=0.944。
  • 观察者内杯尺寸平均ICC=0.965(总体),柄尺寸平均ICC=0.972(总体)。

临床怎么用

  • 原文未据此给出临床推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The purpose of this study is to clarify interobserver and intraobserver reliabilities of the three-dimensional (3D) templating of total hip arthroplasty (THA). METHODS: We selected preoperative computed tomography from 60 hips in 46 patients (14 men and 32 women) who underwent primary THA. To evaluate interobserver and intraobserver reliability, 6 orthopedic surgeons performed 3D templating twice over a 4-week interval. We investigated intraclass correlation coefficients (ICCs) and percent agreement of component size and alignment, comparing morphological differences in the hip. Reproducibility was also compared between groups with osteoarthritis (OA) and those with osteonecrosis (ON). RESULTS: The interobserver reliabilities for mean cup size and stem size were excellent, with ICC = 0.907 and 0.944, respectively. The value was significantly higher in the ON group than in the OA group. In the OA group, the reliability of cup size and alignment decreased in hips with severe subluxation. Percent agreement of stem size was significantly different between the shapes of femoral canal. For intraobserver reliability, the mean ICC of cup size was 0.965 overall, while the value in the ON group was significantly higher than in the OA group. The mean ICC of stem size was 0.972 overall. CONCLUSION: Computed tomography-based 3D templating showed excellent reliability for component size and alignment in THA. Deformity of the affected joint influenced the reliability of preoperative planning.