基于计算机断层扫描的二维到三维匹配导航系统髋臼杯定位的准确性:一项前瞻性随机对照研究

Accuracy of Cup Positioning With the Computed Tomography-Based Two-dimensional to Three-Dimensional Matched Navigation System: A Prospective, Randomized Controlled Study.
PubMed 文献 · J Arthroplasty · 2017-08-12

前瞻性随机对照研究纳入80髋(其中44髋为发育性髋关节发育不良),比较基于CT的2D-3D匹配导航组与配对点匹配导航组的髋臼杯方位准确性;2D-3D组髋臼杯外倾和前倾准确性均优于配对点匹配组,且2D-3D组患者因素对准确性无影响。

要点

  • 2D-3D组髋臼杯外倾准确性为2.5°±2.2°,配对点匹配组为4.6°±3.3°(P=.0016)。
  • 2D-3D组髋臼杯前倾准确性为2.3°±1.7°,配对点匹配组为4.4°±3.3°(P=.0009)。
  • 配对点匹配组中,髋臼顶骨赘存在降低外倾准确性(比值比8.27,P=.0140),骨盆倾斜绝对值对前倾准确性有负面影响(比值比1.27,P=.0222)。

核心结论

  • 基于CT的2D-3D匹配导航的髋臼杯定位准确性优于配对点匹配导航,且不受患者因素影响。
  • 作者认为该系统即使对严重变形的骨盆(如发育性髋关节发育不良)也是有用的系统。

研究设计与样本

  • 前瞻性随机对照研究,共纳入80髋,其中44髋为发育性髋关节发育不良。
  • 分为基于CT的2D-3D匹配导航组(2D-3D组)和配对点匹配导航组(配对点匹配组)。
  • 比较两组髋臼杯方位准确性(术中记录与术后测量的绝对差值),并进行多因素logistic回归分析评估患者因素对准确性的影响。

主要发现

  • 髋臼杯外倾准确性:2D-3D组2.5°±2.2°,配对点匹配组4.6°±3.3°(P=.0016)。
  • 髋臼杯前倾准确性:2D-3D组2.3°±1.7°,配对点匹配组4.4°±3.3°(P=.0009)。
  • 配对点匹配组中,髋臼顶骨赘存在降低外倾准确性(比值比8.27,P=.0140),骨盆倾斜绝对值对前倾准确性有负面影响(比值比1.27,P=.0222)。
  • 2D-3D组中患者因素对髋臼杯方位准确性无影响。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The accuracy of various navigation systems used for total hip arthroplasty has been described, but no publications reported the accuracy of cup orientation in computed tomography (CT)-based 2D-3D (two-dimensional to three-dimensional) matched navigation. METHODS: In a prospective, randomized controlled study, 80 hips including 44 with developmental dysplasia of the hips were divided into a CT-based 2D-3D matched navigation group (2D-3D group) and a paired-point matched navigation group (PPM group). The accuracy of cup orientation (absolute difference between the intraoperative record and the postoperative measurement) was compared between groups. Additionally, multiple logistic regression analysis was performed to evaluate patient factors affecting the accuracy of cup orientation in each navigation. RESULTS: The accuracy of cup inclination was 2.5° ± 2.2° in the 2D-3D group and 4.6° ± 3.3° in the PPM group (P = .0016). The accuracy of cup anteversion was 2.3° ± 1.7° in the 2D-3D group and 4.4° ± 3.3° in the PPM group (P = .0009). In the PPM group, the presence of roof osteophytes decreased the accuracy of cup inclination (odds ratio 8.27, P = .0140) and the absolute value of pelvic tilt had a negative influence on the accuracy of cup anteversion (odds ratio 1.27, P = .0222). In the 2D-3D group, patient factors had no effect on the accuracy of cup orientation. CONCLUSION: The accuracy of cup positioning in CT-based 2D-3D matched navigation was better than in paired-point matched navigation, and was not affected by patient factors. It is a useful system for even severely deformed pelvises such as developmental dysplasia of the hips.