高交联聚乙烯骨水泥型髋臼杯与生物型内衬的股骨头侵入差异
本研究比较了高交联聚乙烯(HXLPE)骨水泥型髋臼杯与生物型内衬在术后5年内的股骨头侵入差异。96例(106髋)平均年龄64.4岁患者分为骨水泥组(35例37髋)和生物型组(61例69髋)。术后5年,两组近端方向股骨头侵入分别为0.103 mm和0.124 mm(P=0.226),前方方向分别为0.090 mm和0.151 mm(P=0.002);平均聚乙烯磨损率在正位片为0.004 mm/y和0.009 mm/y(P=0.286),在Lauenstein位片为0.005 mm/y和0.012 mm/y(P=0.168)。作者结论认为HXLPE使用时所有平均聚乙烯磨损率均很低,骨水泥型髋臼杯与生物型内衬的磨损率非常相似。
要点
- 术后5年,骨水泥组与生物型组近端方向股骨头侵入分别为0.103 mm和0.124 mm(P=0.226),前方方向分别为0.090 mm和0.151 mm(P=0.002)。
- 平均聚乙烯磨损率在正位片为骨水泥组0.004 mm/y vs 生物型组0.009 mm/y(P=0.286),Lauenstein位片为0.005 mm/y vs 0.012 mm/y(P=0.168)。
- 左侧手术和高活动量是正位片聚乙烯磨损的独立危险因素。
核心结论
- 使用HXLPE时,所有平均聚乙烯磨损率均很低,骨水泥型髋臼杯与生物型内衬的磨损率非常相似。
- 聚乙烯颗粒理论提示骨溶解及相关无菌性松动的发生可能因此减少。
研究设计与样本
- 研究纳入96例(106髋)平均年龄64.4岁(范围35-83岁)行全髋关节置换术的患者,分为骨水泥组(35例37髋)和生物型组(61例69髋)。
- 在正位(AP)和Lauenstein位X线片上评估股骨头侵入,基于术后2至5年的股骨头侵入计算平均聚乙烯磨损率。
- 采用多因素分析评估聚乙烯磨损的危险因素。
主要发现
- 术后5年,骨水泥组与生物型组近端方向股骨头侵入分别为0.103 mm和0.124 mm(P=0.226)。
- 术后5年,骨水泥组与生物型组前方方向股骨头侵入分别为0.090 mm和0.151 mm(P=0.002)。
- 平均聚乙烯磨损率在正位片为骨水泥组0.004 mm/y vs 生物型组0.009 mm/y(P=0.286)。
- 平均聚乙烯磨损率在Lauenstein位片为骨水泥组0.005 mm/y vs 生物型组0.012 mm/y(P=0.168)。
- 左侧手术和高活动量是正位片聚乙烯磨损的独立危险因素。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: This study aimed at investigating differences in femoral head penetration between highly cross-linked polyethylene (HXLPE) cemented sockets and uncemented liners during 5 years postoperatively. METHODS: Ninety-six patients (106 hips) with a mean age of 64.4 (range, 35-83) years underwent total hip arthroplasty using a HXLPE cemented socket or liner and were respectively divided into cemented (35 patients [37 hips]) and uncemented (61 patients [69 hips]) groups. Femoral head penetrations were evaluated on both anteroposterior (AP)-view and Lauenstein-view radiographs, and mean polyethylene (PE) wear rates were calculated based on femoral head penetration from 2 to 5 years. Multivariate analyses were performed to assess risk factors for PE wear. RESULTS: At 5 years postoperatively, the cemented and uncemented groups exhibited proximal direction femoral head penetrations of 0.103 mm and 0.124 mm (P = .226) and anterior direction penetrations of 0.090 mm and 0.151 mm (P = .002), respectively. The corresponding mean PE wear rates were 0.004 mm/y and 0.009 mm/y in the AP-view (P = .286) and 0.005 mm/y and 0.012 mm/y in the Lauenstein-view (P = .168), respectively. Left-side operation and high activity were independent risk factors for PE wear on AP-view. CONCLUSION: When HXLPE was used, all mean PE wear rates were very low and those of cemented sockets and uncemented liners were very similar. PE particle theory suggests that the occurrence of osteolysis and related aseptic loosening might consequently decrease.