患者特异性接触应力不能预测特定压配杯的聚乙烯磨损率
本研究纳入45个取自本科室的DuraLoc(DePuy)取出内衬,用液体置换法行三维重量法磨损测定,并用X线片按Blumentritt模型估算髋接触力。结果显示患者特异性因素及估算髋接触力与年磨损率均无相关性,仅VRECAB比值(R=0.408,P=.005)和Alpha角(R=0.377,P=.011)与磨损显著相关,即股骨颈外翻及股骨偏心距较小时磨损增加有限。
要点
- 45个DuraLoc取出内衬中,患者特异性因素与年磨损率无相关性。
- Blumentritt模型估算的髋接触力与年磨损率无相关性。
- VRECAB比值(R=0.408,P=.005)和Alpha角(R=0.377,P=.011)越大磨损越大。
核心结论
- 患者特异性因素及估算髋接触力对DuraLoc杯的年磨损均无主要影响。
- 仅股骨颈外翻和股骨偏心距较小在有限程度上导致磨损增加。
研究设计与样本
- 纳入本科室取出的45个DuraLoc(DePuy)内衬。
- 用液体置换法行三维重量法磨损测定,并用X线片按Blumentritt模型估算髋接触力。
主要发现
- 患者特异性因素与年磨损率无相关性。
- Blumentritt模型估算的髋接触力与年磨损率无相关性。
- VRECAB比值(R=0.408,P=.005)越大磨损越大。
- Alpha角(R=0.377,P=.011)越大(股骨颈越外翻)磨损越大。
关键数值与推荐
- VRECAB比值与年磨损率相关,R=0.408,P=.005。
- Alpha角与年磨损率相关,R=0.377,P=.011。
临床怎么用
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The most common reason for revision total hip arthroplasty remains polyethylene wear. Development dysplasia of the hip and revision situations requires a conscious compromise of implant position. The surgeon should know about the consequence on wear via a possible change in hip contact force. The objective of this study is to investigate whether annual wear is dependent on hip contact force. METHODS: Forty-five inserts (DuraLoc, DePuy) that were explanted in our department were included. Three-dimensional gravimetric determination of the wear was performed by fluid displacement. Then, the hip contact force was determined using radiographs according to the Blumentritt model. RESULTS: No correlation was found between patient-specific factors and the annual wear. The hip contact force estimated by the Blumentritt model also showed no correlation between hip contact force and annual wear. Two single model parameters correlated significantly with wear: VRECAB as a ratio of the lever length of the spinocrural and the pelvitrochanteric muscles and the angle Alpha as a measure of the position of the center of rotation in relation to the greater trochanter. The greater the ratio spinocrural/pelvitrochanteric lever arm (R = 0.408, P = .005) and the greater the Alpha angle (more valgus the femoral neck) (R = 0.377, P = .011) were, the greater was the wear. CONCLUSION: These results lead to the conclusion that neither patient-specific factors nor the estimated hip contact force have a major influence on annual wear in the case of DuraLoc cups. Only a coxa valga and a small femoral offset contribute in a limited amount to an increase in wear.
引用
- Matziolis G, Krakow L, Layher F, et al. Patient-Specific Contact Stress Does Not Predict Polyethylene Wear Rate in a Specific Pressfit Cup. J Arthroplasty. 2017. PMID: 28803815. doi:10.1016/j.arth.2017.07.027
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。