后稳定型全膝关节假体胫骨柱的凸轮柱设计与接触应力:圆角与方角设计的比较
研究用三维有限元模型比较后稳定型全膝关节假体中圆角与方角凸轮柱设计在胫骨柱上的接触应力,发现方角设计应力分布更集中,在屈膝伴10°胫骨内旋及过伸10°时峰值接触应力明显更高,对应接触面积更小。
要点
- 方角凸轮柱设计较圆角设计应力分布更集中。
- 中立位屈膝时两组峰值接触应力相近。
- 屈膝伴10°旋转及过伸10°时方角设计应力明显更高。
核心结论
- 圆角凸轮柱设计在屈膝伴旋转及过伸时比方角设计应力集中更少。
研究设计与样本
- 研究建立两种后稳定型全膝关节假体三维有限元模型(A型与B型),除凸轮柱几何形状外表面相同,A型为圆角凸轮柱设计,B型为方角设计,两者股骨假体髁间窝均为相似弧形。
主要发现
- B型较A型应力分布更集中。
- 中立位屈膝时两组峰值接触应力相近。
- 屈膝伴10°胫骨内旋及过伸10°时B型应力明显高于A型。
- 较高的峰值接触应力对应胫骨柱上较小的接触面积。
关键数值与推荐
- 比较条件为屈膝90°、120°、150°(伴或不伴10°胫骨内旋)及过伸10°时胫骨柱上的应力分布、峰值接触应力与接触面积。
临床怎么用
- 原文未据此给出临床推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The post-cam mechanism in posterior stabilized (PS) prostheses plays an important role in total knee arthroplasty (TKA). The purpose of this study is to clarify the difference of the contact stress on the tibial post between a rounded post-cam design and a squared design during deep knee flexion and at hyperextension using the three-dimensional (3D) finite element models. METHODS: We created 2 types of 3D, finite element models of PS prostheses (types A and B), whose surfaces were identical except for the post-cam geometries: type A has a rounded post-cam design, while type B has a squared design. Both types have a similar curved-shape intercondylar notch of the femoral component. Stress distributions, peak contact stresses, and contact areas on the tibial posts at 90°, 120°, and 150° flexion with/without 10° tibial internal rotation and at 10° hyperextension were compared between the 2 models. RESULTS: Type B demonstrated more concentrated stress distribution compared to type A. The peak contact stresses were similar in both groups during neutral flexion; however, the stresses were much higher in type B during flexion with 10° rotation and at hyperextension. The higher peak contact stresses corresponded to the smaller contact areas in the tibial post. CONCLUSION: A rounded post-cam design demonstrated less stress concentration during flexion with rotation and at hyperextension compared with a squared design. The results would be useful for development of implant designs and prediction of the contact stress on the tibial post in PS total knee arthroplasty.
引用
- Watanabe T, Koga H, Horie M, et al. Post-Cam Design and Contact Stress on Tibial Posts in Posterior-Stabilized Total Knee Prostheses: Comparison Between a Rounded and a Squared Design. J Arthroplasty. 2017. PMID: 28780225. doi:10.1016/j.arth.2017.07.010