金属对聚乙烯全髋关节置换术后因机械辅助缝隙腐蚀行翻修术的早期主要并发症发生率高
本研究纳入27例因单一假体厂商及锥度类型的机械辅助缝隙腐蚀(MACC)行全髋关节翻修术的连续患者,至少随访6个月(平均25个月,范围7-68个月)。翻修术后7例(25.9%)发生主要骨科并发症,6例(22.2%)术后髋关节脱位,3例(11.1%)深部感染,Harris髋关节评分显著改善(P=0.0002),但发生主要并发症者评分较低(70.9 vs 89.2),仅20例(74.1%)获良或优结局。
要点
- 27例MACC翻修患者中7例(25.9%)发生主要骨科并发症
- 术后髋关节脱位6例(22.2%),深部感染3例(11.1%)
- Harris髋关节评分显著改善(P=0.0002),但主要并发症者评分较低(70.9 vs 89.2)
核心结论
- 有症状的MACC是潜在毁灭性诊断,因不良局部组织反应导致软组织和骨活力下降,进而造成早期主要并发症发生率高。
研究设计与样本
- 回顾性队列研究,纳入27例因单一假体厂商及锥度类型的MACC行全髋关节翻修术的连续患者,至少随访6个月(平均25个月,范围7-68个月)。
主要发现
- 主要骨科并发症发生于7/27例(25.9%),均发生于原翻修术后。
- 后续手术中另发生5例主要并发症。
- 术后髋关节脱位发生于6/27例(22.2%)。
- 深部感染发生于3/27例(11.1%)。
- 无菌性髋臼松动、需冲洗和清创的血清肿、肺栓塞、假体周围骨折和再插管各发生于1/27例(3.7%)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Failed total hip arthroplasty caused by mechanically assisted crevice corrosion (MACC) has serious consequences such as adverse local tissue reaction. Revision is currently indicated for significant hip pain, stiffness, and dysfunction; hip instability; progressive bone loss; and soft tissue destruction. Outcomes of this revision surgery are not yet completely understood. METHODS: We examined the surgical outcomes at a minimum of 6 months (average, 25; range, 7-68 months) in a cohort of 27 consecutive patients who underwent revision total hip arthroplasty for MACC associated with a single implant vendor and taper type. RESULTS: Major orthopedic complications occurred in 7 of 27 patients (25.9%), each after the original revision surgery. Five additional major complications occurred in subsequent surgeries. Postoperative hip dislocation occurred in 6 of 27 (22.2%), deep infection occurred in 3 of 27 (11.1%), and aseptic acetabular loosening, seroma requiring irrigation and debridement, pulmonary embolism, periprosthetic fracture, and reintubation each occurred in 1 of 27 (3.7%). Harris hip scores improved significantly with surgery (P = .0002), but overall, scores were lower for those who had major complications (70.9 vs 89.2), and only 20 of 27 patients (74.1%) had good or excellent outcomes. CONCLUSION: Symptomatic MACC is a potentially devastating diagnosis, because of decreased soft tissue and bone viability associated with adverse local tissue reaction. This leads to a high early major complication rate.
引用
- McGrory BJ, Jorgensen AH. High Early Major Complication Rate After Revision for Mechanically Assisted Crevice Corrosion in Metal-on-Polyethylene Total Hip Arthroplasty. J Arthroplasty. 2017. PMID: 28781018. doi:10.1016/j.arth.2017.07.004
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。