手术灯对层流的影响:使用中性浮力氦气泡的实验研究
在模拟全膝关节置换术的层流手术室中,使用中性浮力气泡观察五种不同手术灯配置对层流的影响。结果显示手术灯显著破坏层流并使气泡在术野上方积聚,灯应尽可能远离术野放置。
要点
- 无手术灯时层流存在,气泡快速清除且不积聚。
- 手术灯置于术野上方时层流消失,气泡升至灯后回流至术野。
- 手术灯应尽可能远离术野放置以限制其对层流的负面影响。
核心结论
- 手术灯对层流有显著负面影响,应将灯尽可能远离术野放置以限制此效应。
研究设计与样本
- 在模拟全膝关节置换术的层流手术室中,向术野引入中性浮力气泡,观察气流模式并计数术野上方随时间剩余的气泡数,评估五种不同灯光配置,数据经对数转换后行简单线性回归分析。
主要发现
- 无手术灯时可见层流,气泡快速清除且不积聚。
- 手术灯置于术野上方时层流消失,气泡从术野升至灯后回流至术野。
- 无灯组衰减参数斜率 b=-1.589 与单灯组 b=-0.1273 差异显著(p<0.001)。
- 两灯伸展臂放置组 b=-0.1204 与两灯接触组(p=0.998)及单灯组(p=0.444)效应相似。
关键数值与推荐
- 无灯组斜率 b=-1.589,单灯组 b=-0.1273(p<0.001)。
- 两灯伸展臂组 b=-0.1204 与两灯接触组比较 p=0.998,与单灯组比较 p=0.444。
- 两灯宽距分开160 cm组 b=-1.1165 与无灯组比较 p=0.004。
临床怎么用
- 作者建议手术灯应尽可能远离术野放置以限制其对层流的负面影响。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: The interaction between surgical lighting and laminar airflow is poorly understood. We undertook an experiment to identify any effect contemporary surgical lights have on laminar flow and recommend practical strategies to limit any negative effects. MATERIALS AND METHODS: Neutrally buoyant bubbles were introduced into the surgical field of a simulated setup for a routine total knee arthroplasty in a laminar flow theatre. Patterns of airflow were observed and the number of bubbles remaining above the surgical field over time identified. Five different lighting configurations were assessed. Data were analysed using simple linear regression after logarithmic transformation. RESULTS: In the absence of surgical lights, laminar airflow was observed, bubbles were cleared rapidly and did not accumulate. If lights were placed above the surgical field laminar airflow was abolished and bubbles rose from the surgical field to the lights then circulated back to the surgical field. The value of the decay parameter (slope) of the two setups differed significantly; no light (b = -1.589) versus one light (b = -0.1273, p < 0.001). Two lights touching (b = -0.1191) above the surgical field had a similar effect to that of a single light (p = 0. 2719). Two lights positioned by arms outstretched had a similar effect (b = -0.1204) to two lights touching (p = 0.998) and one light (p = 0.444). When lights were separated widely (160 cm), laminar airflow was observed but the rate of clearance of the bubbles remained slower (b = -1.1165) than with no lights present (p = 0.004). CONCLUSION: Surgical lights have a significantly negative effect on laminar airflow. Lights should be positioned as far away as practicable from the surgical field to limit this effect. Cite this article: Bone Joint J 2017;99-B:1061-6.
引用
- Refaie R, Rushton P, McGovern P, et al. The effect of operating lights on laminar flow: an experimental study using neutrally buoyant helium bubbles. Bone Joint J. 2017. PMID: 28768783. doi:10.1302/0301-620X.99B8.BJJ-2016-0581.R2
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。