后稳定型和限制性髁型全膝关节置换术中聚乙烯损伤随内翻假体对线增加
回顾性分析156个因翻修取出的聚乙烯胫骨垫片,评估假体对线与聚乙烯损伤的关系。术后整体内翻越大,垫片总损伤越重(Spearman相关系数-0.3,95%CI -0.4至-0.1,p=0.001)。翻修前对线较术后更内翻(术后平均外翻3.6°±4.0° vs 翻修前1.7°±6.4°,p=0.04)。
要点
- 术后整体内翻对线越大,取出胫骨垫片的总损伤越重(相关系数-0.3,p=0.001)。
- 翻修前对线较术后向内翻漂移(术后平均外翻3.6° vs 翻修前1.7°,p=0.04)。
- 未发现内侧较外侧聚乙烯损伤增加,可能与假体曲面-曲面环面设计有关。
核心结论
- 尽管手术力求实现中性机械对线,残留内翻对线仍会增加聚乙烯关节面的接触负荷。
- 术后向内翻对线漂移与膝关节内收力矩术后持续较高相一致。
研究设计与样本
- 回顾性研究,纳入2007年至2012年本机构18,065例初次全膝关节置换术中至2016年3月因翻修取出的156个聚乙烯胫骨垫片(来自153例患者),翻修距初次手术均超过2年。
- 排除18例翻修时未取出胫骨垫片者及4例垫片丢失或研究开始前归还患者者。
- 在其他机构行翻修者未纳入,因本研究依赖取回假体。
主要发现
- 术后整体内翻对线更大的膝关节,取出胫骨垫片的总损伤增加(Spearman秩相关系数-0.3,95%CI -0.4至-0.1,p=0.001)。
- 翻修的全膝关节置换术在翻修前倾向于漂回更大内翻,术后对线平均(标准差)为外翻3.6°±4.0°,翻修前为1.7°±6.4°(p=0.04)。
临床怎么用
- 作者指出,尽管手术力求实现中性机械对线,残留内翻对线仍会增加聚乙烯关节面的接触负荷。
局限性
- 本研究为回顾性研究。
- 在其他机构行翻修手术的患者未纳入,因本研究依赖取回假体。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Implant malalignment in primary TKA has been reported to increase stresses placed on the bearing surfaces of implant components. We used a longitudinally maintained registry coupled with an implant retrieval program to consider whether preoperative, postoperative, or prerevision malalignment was associated with increased risk of revision surgery after TKA. QUESTIONS/PURPOSES: (1) What is the relative polyethylene damage on medial and lateral compartments of the tibial plateaus from revised TKAs? (2) Does coronal TKA alignment affect implant performance, such that TKAs aligned in varus are predisposed to experience increased polyethylene damage? (3) Does TKA alignment differ between postoperative and prerevision radiographs, and if so, what does this difference suggest about the mechanical contact load placed on a knee with a TKA? METHODS: Between 2007 and 2012, we performed 18,065 primary TKAs at our institution. By March 2016, 178 of those TKAs (1%) were revised at our center at least 2 years after primary surgery at our institution. Eighteen of those TKAs were excluded from this analysis because the tibial insert was not explanted during revision surgery, and four more were excluded because the inserts were lost or returned to the patient before the study was initiated, leaving 156 retrieved polyethylene tibial inserts (in 153 patients) revised at greater than 2 years after the primary TKA for this retrospective study. Patients who underwent revision surgery elsewhere were not considered here, since this study depended on having retrieved components. Polyethylene damage modes of burnishing, pitting, scratching, delamination, surface deformation, abrasion, and third-body debris were subjectively graded on a scale of 0 to 3 to reflect the extent and severity of each damage mode. On preoperative, postoperative, and prerevision radiographs, overall alignment, femoral alignment, and tibial alignment in the coronal plane were measured according to the protocol recommended by the Knee Society. RESULTS: Knees with more overall varus alignment after TKA had increased total damage on the retrieved tibial inserts (Spearman's rank correlation coefficients of -0.3 [95% CI, -0.4 to -0.1; p = 0.001]). We also found revised TKAs tended to drift back into greater varus before revision surgery, with a mean (SD) of 3.6° ± 4.0° valgus for postoperative alignment compared with 1.7° ± 6.4° prerevision (p = 0.04). CONCLUSIONS: Despite surgical efforts to achieve neutral mechanical alignment, remaining varus alignment places an increased contact load on the polyethylene articular surfaces. The drift toward further varus alignment postoperatively is consistent with the knee adduction moment remaining high after surgery. CLINICAL RELEVANCE: While we found a predisposition toward recurrence of the preoperative varus deformity, we did not find increased medial as opposed to lateral polyethylene damage, which may be explained by the curve-on-curve toroidal design of the articulating surfaces of the TKA implants in this study.
引用
- Li Z, Esposito CI, Koch CN, et al. Polyethylene Damage Increases With Varus Implant Alignment in Posterior-stabilized and Constrained Condylar Knee Arthroplasty. Clin Orthop Relat Res. 2017. PMID: 28822068. doi:10.1007/s11999-017-5477-4
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。