多孔钽单块髋臼组件在初次全髋关节置换术中的生存率:平均随访18年
前瞻性随访128例初次全髋关节置换术(THA)患者(140例中连续纳入),使用压配多孔钽单块髋臼组件,平均随访18.1年(范围17.5-19年)。所有病例Harris髋关节评分、Oxford髋关节评分和关节活动度均显著改善(P < .001),所有髋臼杯影像学稳定,无骨溶解或松动征象;以任何原因再手术为终点的生存率为92.8%,以无菌性松动为终点的生存率为100%。
要点
- 多孔钽单块髋臼杯在初次THA中平均随访18.1年,无菌性松动导致的失败率为0%。
- 以再手术为终点的生存率为92.8%,所有髋臼杯影像学稳定。
- Harris髋关节评分、Oxford髋关节评分和关节活动度在所有病例中显著改善(P < .001)。
核心结论
- 多孔钽单块髋臼杯在初次全髋关节置换术中平均随访18.1年,临床和影像学结果优异,无因无菌性松动导致的失败。
研究设计与样本
- 前瞻性研究,纳入1997年11月至1999年6月间连续128例初次全髋关节置换术(140例患者),所有病例使用压配多孔钽单块髋臼组件,临床和影像学随访平均18.1年(范围17.5-19年)。
主要发现
- 患者手术时平均年龄60.4岁。
- Harris髋关节评分、Oxford髋关节评分和关节活动度在所有病例中均显著改善(P < .001)。
- 末次随访时所有髋臼杯影像学稳定,无迁移、明显聚乙烯磨损、进行性透亮线、骨溶解病变或髋臼骨折的证据。
- 以任何原因再手术为终点的生存率为92.8%,以无菌性松动为终点的生存率为100%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The use of porous tantalum for the acetabular component in primary total hip arthroplasty (THA) has demonstrated excellent short-term and midterm results. However, long-term data are scarce. The purpose of this prospective study is to report the long-term clinical and radiologic outcome following use of an uncemented porous tantalum acetabular component in primary THA with a minimum follow-up of 17.5 years, in a previously studied cohort of patients. METHODS: We prospectively followed 128 consecutive primary THAs in 140 patients, between November 1997 and June 1999. A press-fit porous tantalum monoblock acetabular component was used in all cases. All patients were followed clinically and radiographically for a mean of 18.1 years (range, 17.5-19 years). RESULTS: Mean age of patients at the time of operation was 60.4 years. Harris hip score, Oxford hip score, and range of motion were dramatically improved in all cases (P < .001). At last follow-up, all cups were radiographically stable with no evidence of migration, gross polyethylene wear, progressive radiolucencies, osteolytic lesions, or acetabular fractures. The survivorship with reoperation for any reason as end point was 92.8% and the survivorship for aseptic loosening as an end point was 100%. CONCLUSION: The porous tantalum monoblock cup in primary THA demonstrated excellent clinical and radiographic outcomes with no failures because of aseptic loosening at a mean follow-up of 18.1 years.
引用
- Macheras GA, Lepetsos P, Leonidou AO, et al. Survivorship of a Porous Tantalum Monoblock Acetabular Component in Primary Hip Arthroplasty With a Mean Follow-Up of 18 Years. J Arthroplasty. 2017. PMID: 28734611. doi:10.1016/j.arth.2017.06.049
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。