基于解剖性骨盆平面的骨盆侧卧位定位器联合机械杯导航器在全髋关节置换术中的效能
本研究比较了基于解剖性骨盆平面的骨盆侧卧位定位器联合新型机械杯导航器与传统杯对准导引器在单侧全髋关节置换术中的效果。结果显示,机械导航器组术后杯外倾角和前倾角偏离目标角度的误差均显著小于传统组,表明该导航器能简单、经济、无创地提高侧卧位全髋关节置换术中杯 placement 的准确性。
要点
- 机械导航器组外倾角偏离目标值(2.3°±2.3°)小于传统组(3.5°±3.1°),P=.020。
- 机械导航器组前倾角偏离目标值(3.2°±2.7°)小于传统组(4.6°±3.4°),P=.012。
- 新型导航器可简单、经济、无创地提高侧卧位全髋关节置换术中杯 placement 的准确性。
核心结论
- 可轻松安装于解剖性骨盆平面定位器的机械杯导航器是一种简单、经济、无创的工具,能提高侧卧位全髋关节置换术中杯 placement 的准确性。
研究设计与样本
- 研究纳入136例接受单侧全髋关节置换术并使用解剖性骨盆平面定位器的患者(136髋),传统杯对准导引器组60髋,新型机械杯导航器组76髋。
主要发现
- 传统组平均杯外倾角为39.0°±5.3°,前倾角为21.7°±6.4°;机械导航器组分别为40.6°±3.2°和18.3°±4.6°(P=.018,P<.0001)。
- 传统组杯外倾角偏离目标值为3.5°±3.1°,前倾角偏离目标值为4.6°±3.4°;机械导航器组分别为2.3°±2.3°和3.2°±2.7°(P=.020,P=.012)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The acetabular component orientation in total hip arthroplasty (THA) is of critical importance to the good clinical results. However, traditional widely used cup alignment guides for cup placement are reported to be relatively unreliable. The present study aims at comparing a novel cup alignment guide, which can be attached to our anatomical pelvic plane (APP) pelvic lateral positioner for reducing discrepancies in sagittal pelvic tilt and indicate a targeted cup angle based on the APP, with a conventional cup alignment guide. METHODS: The subjects were 136 hips of 136 patients who underwent unilateral THA using the APP positioner. The procedure was performed with the conventional cup alignment guide (conventional group; 60 hips) and with the novel cup navigator (mechanical navigator group; 76 hips). Postoperative cup angles and discrepancies of postoperative cup angles (inclination and anteversion angles) from the targeted angles were compared between the 2 groups to evaluate the usefulness of these navigators. RESULTS: The mean cup angles in the conventional group were 39.0° ± 5.3° for the inclination angle and 21.7° ± 6.4° for the anteversion angle, whereas those in the mechanical navigator group were 40.6° ± 3.2° and 18.3° ± 4.6°, respectively (P = .018, P < .0001). The discrepancies from the targeted angles were 3.5° ± 3.1° for the inclination angle and 4.6° ± 3.4° for the anteversion angle in the conventional group and 2.3° ± 2.3° and 3.2° ± 2.7°, respectively, in the mechanical navigator group (P = .020, P = .012). CONCLUSION: The mechanical cup navigator easily attachable to the APP positioner is a tool that can improve the accuracy of cup placement in a simple, economical, and noninvasive manner in THA via the lateral position.
引用
- Iwakiri K, Kobayashi A, Ohta Y, et al. Efficacy of a Pelvic Lateral Positioner With a Mechanical Cup Navigator Based on the Anatomical Pelvic Plane in Total Hip Arthroplasty. J Arthroplasty. 2017. PMID: 28712802. doi:10.1016/j.arth.2017.06.033
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。